| 1045 | With G1=152 mbar, G2=156 mbar, P1=16 ubar and mixing chamber temperature of 194.5 mK, attempt to reduce mixing chamber temperature and better understand what issues are affecting the fridge. | |
| 1050 | After reducing V6 to 75%, mixing chamber temperature stabilizes to roughly 194.5 mK. | |
| 1057 | After reducing V6 to 50%, mixing chamber temperature stabilizes to roughly 194.5 mK. | |
| 1105 | After reducing V6 to 25%, P1 has risen to 90 mbar, G1=77.3 mbar adn G2=82.2 mbar, and mixing chamber temperature has risen to 500 mK and is increasing. | |
| 1110 | G1=52.5 mbar, G2=58.6 mbar, P1=1000+mbar and mixing chamber has a temperature of 600 mK and rising. Raise V6 to 40%. | |
| 1118 | G1=272 mbar, G2=230 mbar, P1=.3 mbar, mixing chamber temperature at 382 mK and rising. | |
| 1123 | G1=224 mbar, G2=230 mbar, P1=.3 mbar, mixing chamber temperature at 408 mK and slowly falling. | |
| 1150 | G1=149 mbar, G2=153 mbar, P1=.22 mbar, mixing chamber temperature at 199 mK and oscillating. Changing V6 to 35%. | |
| 1155 | G1=147 mbar, G2=150, P1=.32 mbar, mixing chamber temperature at 205 mK and dropping slightly. | |
| 1205 | G1=152 mbar, G2=154 mbar, P1=.3 mbar, mixing chamber temperature at 224 and dropping. | |
| 1210 | G1=155 mbar, G2=157 mbar, P1=.32 mbar, mixing chamber temperature at 212 and dropping. Change V6 to 45%. Fill LN2 cold trap full. | |
| 1216 | G1=159 mbar, G2=158 mbar, P1=.2 mbar, mixing chamber temperature at 217 and rising. | |
| 1221 | G1=158 mbar, G2=160 mbar, P1=.19 mbar, mixing chamber temperature at 276 and dropping. | |
| 1234 | G1=153 mbar, G2=156 mbar, P1=.2 mbar, mixing chamber temperature at 197 and has started rising. | |
| 1237 | Results of tests suggest restricting flow to still line either increased or does not affect mixing chamber temperature. Commence pumping to dump and stop 1 K Pot. | |
| 1555 | Commence LHe transfer (27.9%). | |
| 1644 | Finish LHe transfer (60.4%). | |
| 1805 | G1=0.8 mbar, G2=733.8 mbar, P1=15 ubar. Close gate valve and condenser speedivalve. Close V6. | |
| 1811 | Commence cryocleaning through coldtrap 1.Fill LN2 coldtrap full. |