2014/05/30
| 1130 | Put up clean tent. |
|---|---|
| 1700 | Uninstall tower. |
| 1800 | Disassemble G106a from tower. Assemble G112 with Pb-210 source and Am-241 source mounted above DIB2. |
| 1815 | Perform tower level electrical check. |
| Detector | G112 | DIB | 1 | |
|---|---|---|---|---|
| side coax | ZB1-4-6 | date/who: | 2014-05-30 | AK & DS |
| face | 2 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | 221 Ω |
| 2 | 4 | QETA-B | 100 to 200 Ω | 246 Ω |
| 3 | 4 | QET B | 100 to 200 Ω | 230 Ω |
| 4 | 6 | QETB-C | 100 to 200 Ω | 231 Ω |
| 5 | 6 | QET C | 100 to 200 Ω | 196 Ω |
| 6 | 8 | QETC-D | 100 to 200 Ω | 217 Ω |
| 7 | 8 | QET D | 100 to 200 Ω | 159 Ω |
| 2 | 8 | QETA-D | 100 to 200 Ω | 224 Ω |
| chassis | 9 | LED1 Rforward | 8 to 9 MΩ | 727 kΩ |
| chassis | 9 | LED1 Vforward | 1.0 to 1.1 V | 1.066 V |
| 9 | chassis | LED1 Rreverse | <1 nS | <6.0 nS |
| chassis | 10 | LED2 Rforward | 8 to 9 MΩ | 650 kΩ |
| chassis | 10 | LED2 Vforward | 1.0 to 1.1 V | 1.063 V |
| 10 | chassis | LED2 Rreverse | <1 nS | closed |
| 1 | 11 | Qo Rbias | 24 to 26 nS | 25 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.2 nS |
| 12 | 13 | Qo Rfeedback | 24 to 26 nS | 26 nS |
| 1 | 14 | Qi Rbias | 24 to 26 nS | 25 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.3 nS |
| 15 | 16 | Qi Rfeedback | 24 to 26 nS | 26 nS |
| Detector | G112 | DIB | 2 | |
|---|---|---|---|---|
| side coax | ZB1-4-8 | date/who: | 2014-05-30 | AK & DS |
| face | 5 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | 187 Ω |
| 2 | 4 | QETA-B | 100 to 200 Ω | 175 Ω |
| 3 | 4 | QET B | 100 to 200 Ω | 109 Ω |
| 4 | 6 | QETB-C | 100 to 200 Ω | 175 Ω |
| 5 | 6 | QET C | 100 to 200 Ω | 146 Ω |
| 6 | 8 | QETC-D | 100 to 200 Ω | 162 Ω |
| 7 | 8 | QET D | 100 to 200 Ω | 107 Ω |
| 2 | 8 | QETA-D | 100 to 200 Ω | 184 Ω |
| chassis | 9 | LED1 Rforward | 8 to 9 MΩ | 735 kΩ |
| chassis | 9 | LED1 Vforward | 1.0 to 1.1 V | 1.060 V |
| 9 | chassis | LED1 Rreverse | <1 nS | <1 nS |
| chassis | 10 | LED2 Rforward | 8 to 9 MΩ | 688 kΩ |
| chassis | 10 | LED2 Vforward | 1.0 to 1.1 V | 1.058 V |
| 10 | chassis | LED2 Rreverse | <1 nS | <8 nS |
| 1 | 11 | Qo Rbias | 24 to 26 nS | 26 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.4 nS |
| 12 | 13 | Qo Rfeedback | 24 to 26 nS | 26 nS |
| 1 | 14 | Qi Rbias | 24 to 26 nS | 26 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.6 nS |
| 15 | 16 | Qi Rfeedback | 24 to 26 nS | 26 nS |
| Detector | G112 | DIB | 3 | |
|---|---|---|---|---|
| side coax | ZB4-4 | date/who: | 2014-05-30 | AK & DS |
| face | 6 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | 178 Ω |
| 2 | 4 | QETA-B | 100 to 200 Ω | 228 Ω |
| 3 | 4 | QET B | 100 to 200 Ω | 210 Ω |
| 4 | 6 | QETB-C | 100 to 200 Ω | 220 Ω |
| 5 | 6 | QET C | 100 to 200 Ω | 197 Ω |
| 6 | 8 | QETC-D | 100 to 200 Ω | 205 Ω |
| 7 | 8 | QET D | 100 to 200 Ω | 148 Ω |
| 2 | 8 | QETA-D | 100 to 200 Ω | 196 Ω |
| chassis | 9 | LED1 Rforward | 8 to 9 MΩ | open |
| chassis | 9 | LED1 Vforward | 1.0 to 1.1 V | open |
| 9 | chassis | LED1 Rreverse | <1 nS | <.1 nS |
| chassis | 10 | LED2 Rforward | 8 to 9 MΩ | open |
| chassis | 10 | LED2 Vforward | 1.0 to 1.1 V | open |
| 10 | chassis | LED2 Rreverse | <1 nS | <.1 nS |
| Nothing unusual found. | |
| 1830 | Fill LN2 coldtrap full. |
| 1940 | Close V5 and V9 to isolate the dump. P1=0.012 mbar, G2= 766 mbar. |
| 1945 | Begin pumping on cold traps with He4 pump. Remove LN2 cold traps from dewar. |