start › cdms › k100 › run_diary › 20140525

2014/05/25

0450 DCRC31 (DIB 2) 15 V power supply seemed to have turned off and on again at some point. Helium transfer was possible cause. As a result, FET had frozen. Turned FET heater on to thaw FET, and turned off again once it had become self-heating. Unfortunately, this meant that the SQUIDs had to be rezapped. Took a few hours for cryostat to cool down again. Restored timing link.
Commence data acquisition with Cf-252 source: 4-hour flash followed by 2-hour cooldown (unsure of how long Side 2 channels were floating, so long flash to be safe). Twelve data series each of length 10 minutes, taken at +/-3.5 V bias, interspersed by 6-second flash and 20-minute cooldown periods. OR trigger on phonon channels.
0530 Fill N2 cold trap full.
1735 Fill N2 cold trap not full: about a litre shy.
1840 Commence data acquisition with Cf-252 source: 30-second flash followed by 45-minute cooldown. Twenty-four data series each of length 10 minutes, taken at +/-3.5 V bias, interspersed by 6-second flash and 20-minute cooldown periods. OR trigger on phonon channels.