UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20140416

2014/04/16

1020 Filled LN2 cold trap full. The village folk rejoice at the safe return of their queen.

Below is a table of the standard deviation of prepulses for the data taken on April 15th over a span of biases. Consistency of data compared for 0 V data, where two series were taken; results of comparison show some small and some larger discrepancies, most notable being the difference of 7 in standard deviation at 0 V for channel 4.

Prepulse Standard Deviation as a Function of Voltage Bias and Channel
Channel 0 V +/-1 V +/-2 V +/-3 V +/-4 V +/-5 V +/-6 V +/-7 V +/-8 V +/-9 V +/-10 V -/+1 V -/+2 V -/+3 V -/+4 V -/+5 V -/+6 V -/+7 V -/+8 V -/+9 V -/+10 V
1 34.0/33.5 37.8 39.0 37.6 39.0 38.3 37.5 34.6 30.0 30.0 23.6 34.5 38.8 40.0 40.1 42.1 45.7 42.5 45.9 48.7 49.0
2 25.8/24.7 27.5 27.6 27.1 28.0 28.1 27.4 26.2 22.1 22.1 15.2 24.6 28.3 29.6 31.1 31.7 34.4 32.8 42.0 36.1 37.8
3 38.2/39.1 46.3 47.6 44.4 46.9 47.4 46.1 44.8 39.1 39.1 32.7 36.6 45.7 46.3 43.7 48.6 51.9 51.3 53.5 60.3 59.1
4 49.2/56.7 59.9 64.3 61.8 62.6 58.9 60.6 56.2 49.0 49.0 42.2 52.5 60.5 61.5 58.0 60.4 62.9 58.2 62.1 63.6 66.7
5 12.1/11.9 13.0 13.8 13.3 13.5 13.8 13.8 13.1 11.2 11.2 8.9 12.4 13.4 13.8 13.9 14.5 15.9 14.5 16.7 15.8 16.7
6 44.0/44.4 50.2 51.1 52.1 52.0 52.1 50.1 50.6 41.1 41.1 37.1 43.2 54.2 54.2 53.7 57.3 60.3 60.1 59.0 66.9 67.8
7 13.8/14.8 17.2 17.4 15.3 17.6 17.7 17.8 18.7 18.2 18.2 17.1 14.0 16.0 16.3 17.5 18.2 20.0 18.8 18.3 21.0 22.4
8 13.6/13.1 13.7 15.0 13.9 13.4 11.7 11.1 9.2 7.5 7.5 6.2 14.4 14.3 15.3 14.2 14.2 13.7 12.9 24.8 12.7 12.9
9 29.4/28.0 33.2 34.5 30.8 33.6 34.1 35.5 35.1 32.5 32.5 29.2 29.0 34.4 36.4 33.1 36.4 38.7 36.9 50.3 43.0 43.1
10 38.5/41.0 46.2 44.9 43.9 46.0 47.0 43.7 43.9 39.6 39.6 32.1 40.1 34.7 45.8 47.8 48.9 53.8 49.8 53.7 57.7 59.4
11 27.9/31.8 32.4 37.3 34.9 38.0 36.5 36.1 33.5 27.9 27.9 22.2 29.8 34.2 35.6 34.4 36.8 39.4 37.5 34.7 40.8 42.8
12 10.4/9.0 9.5 9.5 9.2 8.9 8.0 7.6 7.2 7.0 7.0 6.9 10.4 10.7 10.9 10.2 10.2 28.7 9.7 11.1 9.3 9.1
1950 For completeness, the QET biases used for data collection are reproduced below:
Phonon Channel QET Bias [uA]
DIB1 (DCRC 33) A -64.5
B -79
C -59.5
D -44.5
DIB2 (DCRC 31) A -54
B -68.25
C -60
D -43.5
DIB3 (DCRC 50) A 49
B 65.5
C 40
D 90
2010 Start ionisation data acquisition: 6 hour flash, followed by 2 hour cool down. Trigger on phonon channels, but trace length and lookback time 0.4096 ms & 100 μs respectively. Ba-133 source present. Bias sweeps as follows: +/-12 V; -/+12 V; +/-10 V; & -/+10 V. Data acquisition/flash/cooldown = 5/0.1/20 minutes. Four series at each bias.
2220 Fill N2 cold trap full.