2014/04/14

1100 Fill LN2 cold trap full.
1815 Modify DAQ automation program to unlock and relock SQUIDs after flashing so that higher bias voltages do not prevent the display of pulses.
Start flashing detector for 5 hours followed by 2 hour cooldown and data acquisition. Each bias set will consist of 5-minute data series, interspersed by 6 seconds of flashing and 20 minutes of cooldown. 3.2768 ms trace length; 400 us delay; 200 uS coincidence time; 500 events per file. Data acquired for following biases in order with phonon OR trigger thresholds set at 25 ADC for all phonon channels on each DIB except channels C (DIB2), D (DIB2), and D (DIB3), which were set to thresholds of 40ADC, 30 ADC, and 40ADC, respectively.:
DCRC33 (DIB1) [V] 0 -1 -2 -3 -4 -5 -6 -7 -8 -9 -10 0 1 2 3 4 5 6 7 8 9 10
DCRC31 (DIB2) [V] 0 1 2 3 4 5 6 7 8 9 10 0 -1 -2 -3 -4 -5 -6 -7 -8 -9 -10
DCRC50 (DIB3) [V] 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
1845 Commence helium transfer. Helium liquid level 32.5%.
1915 Cease helium transfer. Helium liquid level 95.8%.
2015 Fill N2 cold trap not full: about a litre shy.