| 1440 | Start flashing detector for 1 minute followed by 20 minute cooldown and data acquisition. DIB1 (DCRC33) set at 0V bias with DIB2 grounded with shorting board for 5-minute data series. 3.2768 ms trace length; 400 us delay; 200 uS coincidence time; 2K events per file. Phonon OR trigger thresholds set at 30 ADC for phonon channels A&D and 35 ADC for phonon channels B&C. |
| 1700 | Performed a rough charge breakdown measurement by biasing the crystal from +/- 12V to -/+ 12V in increments of 1V. With DIB1 set at positive bias voltage and DIB2 set at negative bias, charge breakdown was seen at approximately 9V. |
| | When the crystal was biased with the opposite polarity (i.e. DIB1 and DIB2 at negative and positive biases, respectively) charge breakdown was seen at roughly 8V. |