2014/04/10

1100 Fill LN2 cold trap full.
1230 Continued troubleshooting to try and find the source of the noise seen in the phonon measurements. Took data in the following configurations to see effects:
Attached fan to fridge frame so that it struck the 1K Pot needle valve.
Moved fan away while still turned on so there was no physical contact between blades and needle valve.
Attached copper shielding to DCRC33. (Copper shield could not be attached to DCRC31 without causing issues with powering on.)
Removed all but DCRC33 and shorted DIB2 to ground with shorting board.
2345 Commence helium transfer. Helium liquid level 9.7%.