| 0940 | Stopped flashing detectors. |
|---|---|
| 1325 | Fill LN2 cold trap full. |
| 1700 | Computed QET Bias voltages from IbIs curves. Listed below: |
| DIB 1 | DIB 2 | DIB 3 | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| ChA | ChB | ChC | ChD | ChA | ChB | ChC | ChD | ChA | ChB | ChC | ChD |
| -64.5uA | -79uA | -59.5uA | -44.5uA | -54uA | -68.25uA | -60uA | -43.5uA | -53.5uA | -48uA | -52.5uA | -62.5uA |
| 1810 | Filled LN2 coldtrap full. |
|---|---|
| 1915 | Start flashing detector for 8 hours followed by 2 hour cooldown and data acquisition. Each bias set will consist of three 30-minute data series, interspersed by 30 seconds of flashing and 20 minutes of cooldown. Qi = Qo for each data series. Data acquired for following biases in order: +4 V DCRC31, -4 V DCRC33; -4 V DCRC31, +4 V DCRC33. Trigger thresholds set at DCRC31: 40 Qi, 20 Qo, DCRC33: 35 Qi, 20 Qo. |
| 2100 | DAQDAQ2 has encountered a flashing issue in which the mixing chamber temperature does not increase above 82mK during automated flashing. DAQDAQ2 has been stopped, and manual flashing commenced. |