1100 Fill N2 cold trap full.
1230 Start data acquisition: 2 files at each bias in following order: Q = +12 V, P = 0 V; Q = +12 V, P = +4.8 V; Q = +12 V, P = +12 V; Q = -12 V, P = -12 V; Q = -12 V, P = -4.8 V; Q = -12 V, P = 0 V; . Initial flash for 20 minutes, followed by 1-hour cooldown; acquisition/flash/cooldown = 5/0.1/20 minutes.
2330 Fill N2 cold trap full.
2350 Commence helium transfer. Helium liquid level 12.6%.
Table 1: Positions of 356 keV peak in overnight Q = -4 V, P = 0 V data series.
Data Series Q Only Mean Q + P Mean
07140205_2210 0.01150 ± 0.00002 0.01254 ± 0.00002
07140205_2310 0.01145 ± 0.00002 0.01248 ± 0.00003
07140206_0011 0.01148 ± 0.00001 0.01251 ± 0.00002
07140206_0111 0.01146 ± 0.00002 0.01251 ± 0.00002
Mean 0.01147 0.01251
Variance 0.00005 0.00006
Standard Deviation 0.00002 0.00002