| 1355 | Filled LN2 coldtrap full. |
| 1545 | Performed single shot test. Tower thermometer reached a minimum temperature of ~51 mK during test. |
| 1550 | Commence pumping mixture to dump. Turn off 1K Pot. |
| 1650 | Commence He transfer. Helium level at 23.0%. |
| 1740 | End LHe transfer (94.9%.) |
| 2000 | Mixing chamber temperature >6 K. Dump pressure has remained at 669 mbar for the last couple of hours, so most of the mixture has been recovered. Turn off mixing chamber heater, stop recovering mixture and start 1 K pot (temperature 1.82 K). |
| | Fill N2 cold trap full. Commence manual condensation procedure through nitrogen cold trap 1. |
| 2030 | Valve V12A fully open: manual condensation procedure complete. Dump pressure approximately 319 mbar (350 mbar dump pressure of mixture in fridge). Mixing chamber temperature 5.1 K. |
| 2245 | Mixing chamber temperature 2.6 K. Commence manual circulation procedure. |
| 2350 | Valve V6 fully open. Cease manual circulation procedure. |