| 1040 | Filled LN2 coldtrap full. Fridge temp at 59.3 mK. |
| 1450 | Tower thermometer reads 72.1 mK (mixing chamber temperature 59.2 mK), which implies quite a large thermal gradient. According to Nader, a temperature gradient of around 10 mK is normal considering the number of interfaces between the mixing chamber and tower. |
| 1615 | Plugged DCRC31 into DIB 2, and noise is still present as with previous runs at same level of aroudn 300 mV peak-to-peak. Interestingly, tower and mixing chamber temperatures rose to around 124 mK and 75 mK respectively. The rise in mixing chamber temperature was not witnessed during previous runs. Would be interesting to see the effect of plugging DCRC31 into DIB 1. |
| 1705 | Plugged DCRC31 into DIB 1, signals look normal. Tower and mixing chamber temperatures stabilized to 72.1 mK and 59.3 mK respectively, thus no rise in temperature was seen. |
| 2355 | Fill N2 cold trap full. |