2013/12/05
| 1443 | Perform electrical check on G114. Housing number for G114 is PF-235-933-91 Rev2. |
|---|
| Detector | G114 | DIB | 1 | |
|---|---|---|---|---|
| date/who: | 2013-12-05 | DS AK | ||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | open |
| 3 | 4 | QET B | 100 to 200 Ω | 167Ω |
| 5 | 6 | QET C | 100 to 200 Ω | 130Ω |
| 7 | 8 | QET D | 100 to 200 Ω | 72Ω |
| 9 | chassis | LED1 Rforward | 8 to 9 MΩ | open |
| 9 | chassis | LED1 Vforward | 1.0 to 1.1 V | open |
| chassis | 9 | LED1 Rreverse | <1 nS | < .1 nS |
| 10 | chassis | LED2 Rforward | 8 to 9 MΩ | 783kΩ |
| 10 | chassis | LED2 Vforward | 1.0 to 1.1 V | 1.042V |
| chassis | 10 | LED2 Rreverse | <1 nS | <0.1 nS |
| 11 | 12 | Qo-Qi | 87Ω | |
| 12 | 11 | Qi-Qo | 83Ω |
| Detector | G114 | DIB | 2 | |
|---|---|---|---|---|
| date/who: | 2013-12-05 | DS AK | ||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | 155Ω |
| 3 | 4 | QET B | 100 to 200 Ω | 35Ω |
| 5 | 6 | QET C | 100 to 200 Ω | 108Ω |
| 7 | 8 | QET D | 100 to 200 Ω | 35Ω |
| 9 | chassis | LED1 Rforward | 8 to 9 MΩ | 755kΩ |
| 9 | chassis | LED1 Vforward | 1.0 to 1.1 V | 1.042V |
| chassis | 9 | LED1 Rreverse | <1 nS | < .1 nS |
| 10 | chassis | LED2 Rforward | 8 to 9 MΩ | 754kΩ |
| 10 | chassis | LED2 Vforward | 1.0 to 1.1 V | 1.044V |
| chassis | 10 | LED2 Rreverse | <1 nS | < 1 nS |
| 11 | 12 | Qo-Qi | 194kΩ | |
| 12 | 11 | Qi-Qo | 1.2MΩ |
| Detector | G114 | DIB | 3 | |
|---|---|---|---|---|
| date/who: | 2013-12-05 | DS AK | ||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | 31Ω |
| 3 | 4 | QET B | 100 to 200 Ω | 81Ω |
| 5 | 6 | QET C | 100 to 200 Ω | 30Ω |
| 7 | 8 | QET D | 100 to 200 Ω | 83Ω |
| 9 | chassis | LED1 Rforward | 8 to 9 MΩ | open |
| 9 | chassis | LED1 Vforward | 1.0 to 1.1 V | open |
| chassis | 9 | LED1 Rreverse | <1 nS | open |
| 10 | chassis | LED2 Rforward | 8 to 9 MΩ | open |
| 10 | chassis | LED2 Vforward | 1.0 to 1.1 V | open |
| chassis | 10 | LED2 Rreverse | <1 nS | open |
| 11 | 12 | Qo-Qi | open | |
| 12 | 11 | Qi-Qo | open |
| 1620 | Mount G101b on tower with Am source facing side 2. Perform electrical check. |
|---|
| Detector | G101b | DIB | 1 | |
|---|---|---|---|---|
| side coax | ZB1-1-9 | date/who: | 2013-12-05 | DS AK |
| face | 2 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | open |
| 3 | 4 | QET B | 100 to 200 Ω | open |
| 5 | 6 | QET C | 100 to 200 Ω | open |
| 7 | 8 | QET D | 100 to 200 Ω | open |
| 9 | chassis | LED1 Rforward | 8 to 9 MΩ | 762 kΩ |
| 9 | chassis | LED1 Vforward | 1.0 to 1.1 V | 1.103 V |
| chassis | 9 | LED1 Rreverse | <1 nS | < 2 nS |
| 10 | chassis | LED2 Rforward | 8 to 9 MΩ | open |
| 10 | chassis | LED2 Vforward | 1.0 to 1.1 V | open |
| chassis | 10 | LED2 Rreverse | <1 nS | <0.6 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.5 nS |
| 12 | 13 | Qo Rfeedback | 24 to 26 nS | 26 nS |
| 11 | 14 | Q Bias | 12.5 nS | 13 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.5 nS |
| 15 | 16 | Qi Rfeedback | 24 to 26 nS | 26 nS |
| Detector | G101b | DIB | 2 | |
|---|---|---|---|---|
| side coax | ZB1-6 | date/who: | 2013-12-05 | DS AK |
| face | 5 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | open |
| 3 | 4 | QET B | 100 to 200 Ω | open |
| 5 | 6 | QET C | 100 to 200 Ω | open |
| 7 | 8 | QET D | 100 to 200 Ω | open |
| 9 | chassis | LED1 Rforward | 8 to 9 MΩ | 781 kΩ |
| 9 | chassis | LED1 Vforward | 1.0 to 1.1 V | 1.104 V |
| chassis | 9 | LED1 Rreverse | <1 nS | <0.5 nS |
| 10 | chassis | LED2 Rforward | 8 to 9 MΩ | 781 kΩ |
| 10 | chassis | LED2 Vforward | 1.0 to 1.1 V | 1.108 V |
| chassis | 10 | LED2 Rreverse | <1 nS | <1 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.25 nS |
| 12 | 13 | Qo Rfeedback | 24 to 26 nS | 26 nS |
| 11 | 14 | Q Bias | 12.5 nS | 13 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.3 nS |
| 15 | 16 | Qi Rfeedback | 24 to 26 nS | 26 nS |
| 1820 | Mount tower in IVC. Dib 1 is on side 2 of the tower and is pointing towards the center of lab (between striplines and 600 mk rod. Dib 2 is on side 5 and is point towards the corner. |
|---|