| 0345 | Seal up and begin pumping on IVC with rough pump. |
| 0600 | IVC pressure 5.4E-4 m bar. Open line to leak checker and isolate turbo pump. Background helium leak rate 1.5E-7 mbar l/s. Evacuate main bath and overpressurise with He-4 gas. Helium leak rate decreased to 1.4E-7 mbar l/s. |
| | Fill N2 cold trap full. |
| 0615 | Evacuate 1 K pot. |
| | Add 1 mbar He-4 exchange gas to IVC. Commence nitrogen transfer. |
| 0810 | Cease nitrogen transfer. Mixing chamber temperature 140 K and falling, IVC pressure 0.46 mbar. |
| 1105 | Mixing chamber temperature 82 K. Start evacuating IVC. |
| 1340 | IVC pressure 2.5E-6 mbar. |
| 1347 | Fill LN2 cold trap full. |
| 1400 | With mixing chamber temperature of 81.5 K commence Nitrogen blowout. Simultaneously commence leak check with background leak rate of 2.4E-7 mbar*l/s and IVC pressure of 2.2E-5 mbar. |
| 1505 | End Nitrogen blow out. Leak rate dropped consistently during course of transfer, ending with 1.4E-7 mbar*l/s. There is no indication of a leak. Mixing chamber temperature at 81.2 K and IVC pressure of 3.5E-5 mbar. |
| 1510 | Begin pumping on 1K Pot and evacuating main bath. |
| 1530 | Start leak check with background leak rate at 1.3E-7 mbar*l/s. Begin filling main bath with He exchange gas. |
| 1541 | Main bath filled with exchange gas, leak rate held at 1.2E-7 mbar*l/s. As before, no sign of leak. |
| | Switch to turbo pump and turn off leak checker. |
| 1605 | Add 40 μbar He exchange gas. |
| 1620 | Commence LHe transfer (0.0%). |
| 1920 | End LHe transfer (95.0%). |
| 2100 | With IVC pressure of 6.3E-3 mbar and mixing chamber temperature of 10.9 K, commence pumping out IVC exchange gas with Turbo pump. |