1210 Filled LN2 cold trap full.
Detector G114 DIB 1
date/who: 2013-11-11 DS AK
+lead - lead signal typical measured
1 2 QET A 100 to 200 Ω open
3 4 QET B 100 to 200 Ω 206Ω
5 6 QET C 100 to 200 Ω 140Ω
7 8 QET D 100 to 200 Ω 72Ω
chassis 9 LED1 Rforward 8 to 9 MΩ open
chassis 9 LED1 Vforward 1.0 to 1.1 V open
9 chassis LED1 Rreverse <1 nS < .1 nS
chassis 10 LED2 Rforward 8 to 9 MΩ 783kΩ
chassis 10 LED2 Vforward 1.0 to 1.1 V 1.057V
10 chassis LED2 Rreverse <1 nS <0.1 nS
11 12 Qo-Qi 114Ω
12 11 Qi-Qo 110Ω
Detector G114 DIB 2
date/who: 2013-11-11 DS AK
+lead - lead signal typical measured
1 2 QET A 100 to 200 Ω 191Ω
3 4 QET B 100 to 200 Ω 35Ω
5 6 QET C 100 to 200 Ω 114Ω
7 8 QET D 100 to 200 Ω 35Ω
chassis 9 LED1 Rforward 8 to 9 MΩ 778kΩ
chassis 9 LED1 Vforward 1.0 to 1.1 V 1.055V
9 chassis LED1 Rreverse <1 nS < .1 nS
chassis 10 LED2 Rforward 8 to 9 MΩ 781kΩ
chassis 10 LED2 Vforward 1.0 to 1.1 V 1.058V
10 chassis LED2 Rreverse <1 nS < 1 nS
11 12 Qo-Qi 465kΩ
12 11 Qi-Qo 2.8MΩ
Detector G114 DIB 3
date/who: 2013-11-11 DS AK
+lead - lead signal typical measured
1 2 QET A 100 to 200 Ω 32Ω
3 4 QET B 100 to 200 Ω 134Ω
5 6 QET C 100 to 200 Ω 32Ω
7 8 QET D 100 to 200 Ω 85Ω
chassis 9 LED1 Rforward 8 to 9 MΩ open
chassis 9 LED1 Vforward 1.0 to 1.1 V open
9 chassis LED1 Rreverse <1 nS open
chassis 10 LED2 Rforward 8 to 9 MΩ open
chassis 10 LED2 Vforward 1.0 to 1.1 V open
10 chassis LED2 Rreverse <1 nS open
11 12 Qo-Qi open
12 11 Qi-Qo open
1800 After detector checkout, detector was placed in second cabinet from top.
Lead source was moved to bottom shelf.
2140 Assume Vq (charge lines) are connected to pin 12 (Qi) and Vp (phonon lines) are connected to pin 11 (Qo). Started with Vp = Vq = -4 V, and adjusted Vp in 0.5 V steps towards 0 V. Although asymmetric collection was first witnessed at Vp = -3 V, at no point did symmetric collection completely disappear. Perhaps this is expected. Instead, let's use Vp = b*Vq, where b = 0.4. This value was found by Scott Hertel to be the maximum boost for 25 mm thick crystals before bulk discrimination was compromised.
Noticed that default 0 V bias values for DCRC31 were incorrect. Not sure how that happened. Set new defaults.
Start flashing detector for 5 hours followed by 2 hour cooldown and data acquisition. Each bias set will consist of four 10-minute data series, interspersed by 30 seconds of flashing and 20 minutes of cooldown. Data will be acquired in following order:
Vp [V] Vq [V]
-12 -4.8
-8 -3.2
-4 -1.6
-2 -0.8
-12 0.0
-8 0.0
-4 0.0
-2 0.0