2013/10/23
| 1330 | Pressures in P1 & G1 are 16 μbar & 1 mbar respectively. Isolate turbo pump. |
|---|---|
| 2010 | Pressures in P1 & G1 are 20 μbar & 1.3 mbar respectively. Recommence evacuation of mixing chamber lines with turbo pump. |
| Detector | G101b | DIB | 1 | |
|---|---|---|---|---|
| side coax | ZB1-1-9 | date/who: | 2013-10-23 | HR AK |
| face | 2 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | open |
| 3 | 4 | QET B | 100 to 200 Ω | open |
| 5 | 6 | QET C | 100 to 200 Ω | open |
| 7 | 8 | QET D | 100 to 200 Ω | open |
| chassis | 9 | LED1 Rforward | 8 to 9 MΩ | 766 kΩ |
| chassis | 9 | LED1 Vforward | 1.0 to 1.1 V | 1.103 V |
| 9 | chassis | LED1 Rreverse | <1 nS | < 1 nS |
| chassis | 10 | LED2 Rforward | 8 to 9 MΩ | open |
| chassis | 10 | LED2 Vforward | 1.0 to 1.1 V | open |
| 10 | chassis | LED2 Rreverse | <1 nS | <0.2 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.2 nS |
| 12 | 13 | Qo Rfeedback | 24 to 26 nS | 26 nS |
| 11 | 14 | Q Bias | 12.5 nS | 13 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.2 nS |
| 15 | 16 | Qi Rfeedback | 24 to 26 nS | 26 nS |
| Detector | G101b | DIB | 2 | |
|---|---|---|---|---|
| side coax | ZB1-6 | date/who: | 2013-10-23 | HR AK |
| face | 5 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | open |
| 3 | 4 | QET B | 100 to 200 Ω | open |
| 5 | 6 | QET C | 100 to 200 Ω | open |
| 7 | 8 | QET D | 100 to 200 Ω | open |
| chassis | 9 | LED1 Rforward | 8 to 9 MΩ | 734 kΩ |
| chassis | 9 | LED1 Vforward | 1.0 to 1.1 V | 1.105 V |
| 9 | chassis | LED1 Rreverse | <1 nS | <0.5 nS |
| chassis | 10 | LED2 Rforward | 8 to 9 MΩ | 783 kΩ |
| chassis | 10 | LED2 Vforward | 1.0 to 1.1 V | 1.108 V |
| 10 | chassis | LED2 Rreverse | <1 nS | <0.4 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.25 nS |
| 12 | 13 | Qo Rfeedback | 24 to 26 nS | 26 nS |
| 11 | 14 | Q Bias | 12.5 nS | 13 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.3 nS |
| 15 | 16 | Qi Rfeedback | 24 to 26 nS | 26 nS |