UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20130828

2013/08/28

1115 Filled LN2 coldtrap.
1145 Begin LHe transfer (-1.0%)
1205 Triggers set at 26 (Qi DIB 1), 40 (Qo DIB 1), 36 (Qi DIB 2) & 40 (Qo DIB 2) ADC bins. Commence with 4-hour flash followed by 2-hour cooldown. Four files (1000 events per file) at each bias in 5/0.1/20 minutes acquisition/flash/cooldown sequence. Order of biases: +/-5 V; -/+5 V; +/-6 V;-/+6 V; +/-10 V; -/+10 V; .
1250 End LHe transfer (89.8%)
2200 Interrupt data acquisition during second +/-6 V data series 07130828_2157. Commence LED tests. Reasons outlined in section below:

Questionable Detector Neutralisation

Paul raised the point about neutrlisation during Monday's DTC meeting, and it seems as though the crystal was not neutralised sufficiently well at the beginning of this run. Fig. 1 shows Qsum spectra from the first four data series at -/+4 V bias, taken after the first 4-hour flash and +/-4 V data series. The peak on Side 1 appears to be pretty clear, and a lot of electrons are lost on their way to Side 2 resulting in a peak at roughly 1.1 mV. However, Qsum spectra data series at the same bias a couple of days later after another 4-hour flash indicate that the Side 2 peak has shifted to a mean position of 1.6 mV, as shown in Fig. 2. The only explanation for this is poor neutralisation, thus falsely implying poor charge collection efficiency for electrons at this bias.

Figure 1: Qsum Spectra for Sides 1 & 2; -/+4 V bias; taken on 25th August 2013.
Figure 2: Qsum Spectra for Sides 1 & 2; -/+4 V bias; taken on 27th August 2013.

The situation gets worse, as the following day data series were acquired at -/+3.5 V (Fig. 3) & -/+4.5 V (Fig. 4) biases after another 4-hour flash. The 60 keV peak on Side 2 in Fig. 3 has a mean position just shy of 1.8 mV, which would imply that charge collection efficiency is greater at a lower bias. Of course, this doesn't make sense, and would imply that the detector was still not fully neutralised.

Figure 3: Qsum Spectra for Sides 1 & 2; -/+3.5 V bias; taken on 28th August 2013.
Figure 4: Qsum Spectra for Sides 1 & 2; -/+4.5 V bias; taken on 28th August 2013.

Tests should be performed on the LEDs to find an ideal flashing scheme. Additionally, data at a few select biases should be taken after a long flash, and compared with the data retaken at these biases after another long flash to determine if the new scheme has worked.