| 0150 | G2 pressure 684 mbar: mixture recovered. Turn off mixing chamber heater. |
| 0215 | Commence helium transfer. Helium liquid level 10.1%. |
| 0235 | Fill N2 cold trap full. |
| 0240 | Close gate valve and condenser line speedivalve to isolate fridge. Close nitrogen cold trap 1 speedivalve and all valves on IGH. Commence sryocleaning mixture through nitrogen cold trap 2. |
| 0305 | Cease helium transfer. Helium liquid level 90.9%. |
| 0410 | Pump mixture into dump in preparation for condensation. |
| 0430 | Start 1 K pot. |
| 0440 | Temperature of 1 K pot 1.88 K. Open nitrogen cold trap 1 speedivalve. Open gate valve and condenser line speedivalve. Mixture in dump 689 mbar. Start manual condensation procedure through nitrogen cold trap 1. |
| 0530 | Valve V12A fully open: manual condensation procedure complete. Mixture left in dump 364 mbar: approximately 325 mbar of mixture in fridge. |
| 0600 | Reduce temperature of 1 K pot to 1.73 K. |
| 1800 | Fill N2 cold trap full. |
| | Mixing chamber temperature 1.81 K. Temperature of 1 K pot 1.72 K and steady. Pressures in G1 & G2 at around 40-50 mbar. Mixture has condensed. Commence manual circulation procedure. |
| 1820 | Valve V6 fully open. Manual circulation procedure complete. Wait for mixing chamber to reach base temperature. |