2013/08/19
| 1130 | Filled LN2 cold trap. |
|---|---|
| 1440 | Pump mixture into dump in preparation for condensation. |
| 1600 | Mixing chamber temperature 8.1K. IVC pressure 4.0e-6 mbar. |
| Isolate IVC and start 1K pot. | |
| 1630 | 1K pot temperature 1.8K. Open condenser speedivalve and gate valve. Pressure in dump 690 mbar. |
| Commence manual condensation procedure through nitrogen cold trap 1. | |
| 1700 | 339 mbar of mixture left in dump. |
| 1735 | Cease manual condensation procedure: valve V12A on IGH fully open. Wait for mixing chamber temperature to decrease further. |
| 1920 | Mixing temperature temperature between 4.2K and 3.1K during electrical check. |
| QET and QET Return lines seem to have a slight short to ground. Found conductances that decrease to 1.0nS-2.5nS on various QET lines. However, we've decided to assess this problem again at base temperature. |
| Detector: | G106a DIB1 | BH: | Stripline: | 27 | ||||||||
| SC: | ZB1-4-4 | 300 K | 80K | LHe | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 2 | Date | 2013-08-14 | Date | 2013-08-17 | Date | 2013-08-17 | |||||
| SQUET: | ZC-70 | Who | CP | Who | CP | Who | CP | |||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis | |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | ||||||||||
| Nominal: | Nominal: | Nominal: | ||||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | ||||||||||
| Phonon | ||||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.8k | OK | 1 - 2 k | 1.9k | OK | 80 - 140 | 100 | OK | |
| SQUID bias B/Ret | 19 | 34 | 4.4k | OK | 1.7k | OK | 80 - 140 | 105 | OK | |||
| SQUID bias C/Ret | 8 | 24 | 4.6k | OK | 1.8k | OK | 80 - 140 | 97 | OK | |||
| SQUID bias D/Ret | 23 | 39 | 5.2k | OK | 2.1k | OK | 80 - 140 | 96 | OK | |||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 29k | OK | 9 - 13 k | 15k | OK | 1.0 - 1.3 k | 1.2k | OK | |
| Feedback B/Ret | 2 | 18 | 25k | OK | 9.2k | OK | 1.0 - 1.3 k | 1.2k | OK | |||
| Feedback C/Ret | 41 | 25 | 28k | OK | 14k | OK | 1.0 - 1.3 k | 1.3k | OK | |||
| Feedback D/Ret | 22 | 6 | 34k | OK | 17k | OK | 1.0 - 1.3 k | 1.2k | OK | |||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 112, 111, 91, 93, open | 1.3nS | 30 - 40 | 74, 64, 53, 56, open | OK | 20 - 30 | 55, 46, 32, 35, open | 1.6nS | |
| QET B/Ret | 35 | 29, 33, 47, 48 | 80, 79, 114, 79 | 1.3nS | 41, 40, 80, 40 | OK | 20 - 30 | 26, 25, 46, 25 | 1.7nS | |||
| QET C/Ret | 7 | 5, 9, 10, 28 | open, 81, 72, 80 | 1.3nS | open, 45, 36, 44 | OK | 20 - 30 | open, 26, 17, 25 | 1.3nS | |||
| QET D/Ret | 38 | 36, 40, 43, 44 | 85, open, 83, 83 | 1.3nS | 48, open, 45, 45 | OK | 20 - 30 | 27, open, 25, 25 | 1.4nS | |||
| Charge | ||||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.06k | OK | 1.0 - 1.1 k | 1.04k | OK | 1.0 - 1.1 k | 1.2k | OK | |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 26 | 1.1nS | < 1 nS | <0.1ns | OK | <1 nS | <0.3nS | OK | ||
| Qo bias/QET Ret | 11 | 26 | 1.1nS | <0.1ns | OK | <1 nS | <0.3nS | OK | ||||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 24, 24 | OK | 6 - 8 nS | 21, 21 | OK | <1 nS | 19, 19 | OK | |
| Qo feedback/Drain/Source | 30 | 12, 27 | 24, 24 | OK | 21, 21 | OK | <1 nS | 20, 20 | OK | |||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK | |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | open | OK | 0.9 - 1.0 V | 1.08 | OK | — | – | OK | |
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.12 | – | 1.2 - 1.5 V | 1.4 | – | — | – | – | |
| LED 2 | chassis | 26 | 1.12 | – | 1.4 | – | — | – | – | |||
| Detector: | G106a DIB2 | BH: | Stripline: | 19 | |||||||
| SC: | ZB1-4-6 | 300 K | LN | LHe | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 5 | Date | 2013-08-14 | Date | 2013-08-17 | Date | 2013-08-19 | ||||
| SQUET: | ZC-13 | Who | CP | Who | CP | Who | CP | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.6k | OK | 1 - 2 k | 1.7k | OK | 80 - 140 | 109 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.8k | OK | 1.8k | OK | 80 - 140 | 104 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 5.3k | OK | 2.1k | OK | 80 - 140 | 123 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 4.9k | OK | 1.9k | OK | 80 - 140 | 103 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 18k | OK | 9 - 13 k | 7.5k | OK | 1.0 - 1.3 k | 1.3k | OK |
| Feedback B/Ret | 2 | 18 | 20k | OK | 8.1k | OK | 1.0 - 1.3 k | 1.3k | OK | ||
| Feedback C/Ret | 41 | 25 | 29k | OK | 11.6k | OK | 1.0 - 1.3 k | 1.3k | OK | ||
| Feedback D/Ret | 22 | 6 | 21k | OK | 8.5k | OK | 1.0 - 1.3 k | 1.2k | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 124, 108, 92, 94, 93 | 1.2 | 30 - 40 | 77, 70, 55, 57, 56 | OK | 20 - 30 | 62, 52, 40, 43, 41 | 2.5nS |
| QET B/Ret | 35 | 29, 33, 47, 48 | 87, 86, 110, 86 | 1.2 | 48, 47, 77, 47 | OK | 20 - 30 | 32, 31, 60, 31 | 1.9nS | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 102, 90, 79, 89 | 1.2 | 60, 48, 37, 47 | OK | 20 - 30 | 46, 33, 22, 31 | 0.8nS | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 90, 87, 87, 87 | 1.3 | 50, 47, 47, 47 | OK | 20 - 30 | 33, 31, 30, 31 | 1.6nS | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.08k | OK | 1.0 - 1.1 k | 1.01k | OK | 1.0 - 1.1 k | 1.23k | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 25 | OK | < 1 nS | <0.2nS | OK | <1 nS | <0.5ns | OK | |
| Qo bias/QET Ret | 11 | 26 | OK | <0.1nS | OK | <1 nS | <0.4ns | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23, 23 | OK | 6 - 8 nS | 20, 20 | OK | <1 nS | 19, 19 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23, 23 | OK | 20, 20 | OK | <1 nS | 19, 19 | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.7 | OK | 0.9 - 1.0 V | 1.09 | OK | — | – | OK |
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.12 | – | 1.2 - 1.5 V | 1.41 | – | — | – | – |
| LED 2 | chassis | 26 | 1.12 | – | 1.39 | – | — | – | – | ||
| Detector: | G106a DIB3 | BH: | Stripline: | 20 | |||||||
| SC: | ZB1-4-8 | 300 K | LN | LHe | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 6 | Date | 2013-08-14 | Date | 2013-08-17 | Date | 2013-08-19 | ||||
| SQUET: | ZC-23 | Who | CP | Who | CP | Who | CP | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4k | OK | 1 - 2 k | 1.6k | OK | 80 - 140 | 98 | OK |
| SQUID bias B/Ret | 19 | 34 | 5.0k | OK | 2.0k | OK | 80 - 140 | 96 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.6k | OK | 1.7k | OK | 80 - 140 | 96 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 4.3k | OK | 1.6k | OK | 80 - 140 | 90 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 18k | OK | 9 - 13 k | 7.3k | OK | 1.0 - 1.3 k | 1.3k | OK |
| Feedback B/Ret | 2 | 18 | 20k | OK | 8.2k | OK | 1.0 - 1.3 k | 1.3k | OK | ||
| Feedback C/Ret | 41 | 25 | 19k | OK | 7.6k | OK | 1.0 - 1.3 k | 1.2k | OK | ||
| Feedback D/Ret | 22 | 6 | 18k | OK | 7.0k | OK | 1.0 - 1.3 k | 1.2k | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 107, 93, 77, 79, 78 | 1.3nS | 30 - 40 | 70, 62, 46, 48, 46 | OK | 20 - 30 | 54, 47, 31, 33, 31 | 1.9nS |
| QET B/Ret | 35 | 29, 33, 47, 48 | 72, 72, 91, 72 | 1.3nS | 39, 38, 69, 38 | OK | 20 - 30 | 25, 24, 56, 24 | 2.0nS | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 81, 71, 63, 70 | 1.3nS | 53, 37, 29, 36 | OK | 20 - 30 | 34, 24, 16, 23 | 1.8nS | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 73, 71, 70, 71 | 1.3nS | 39, 37, 37, 37 | OK | 20 - 30 | 26, 24, 24, 24 | 2.1nS | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.06k | OK | 1.0 - 1.1 k | 1.04k | OK | 1.0 - 1.1 k | 1.2k | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 0.1nS | OK | < 1 nS | <0.1nS | OK | <1 nS | <0.4nS | OK | |
| Qo bias/QET Ret | 11 | 0.1nS | OK | <0.1nS | OK | <1 nS | <0.3nS | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23, 23 | OK | 6 - 8 nS | 20, 20 | OK | <1 nS | 18, 18 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23, 23 | OK | 20, 20 | OK | <1 nS | 19, 19 | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.72 | OK | 0.9 - 1.0 V | 1.09 | OK | — | – | OK |
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.1 | 10 nS | 1.2 - 1.5 V | 1.4 | OK | — | – | – |
| LED 2 | chassis | 26 | 1.1 | open | 1.39 | 23 nS | — | – | 13 nS | ||
| 2205 | Switch thermometer. Mixing chamber temperature 2.2 K and pressures in G1 & G2 are less than 70 mbar. Close valve V3 on IGH and commence manual circulation procedure. |
|---|---|
| 2215 | Fill N2 cold trap full. |
| 2230 | Valve V6 fully open on IGH; cease manual circulation procedure. Await mixing chamber to reach base temperature. |