| 0930 | Temperature has decreased and mixture is has been condensing. Lakeshore/MC=30.2K, G1=188mbar, G2=525mbar. |
| 1617 | Started helium transfer. Main bath at 17.1%. |
| 1642 | Finished helium transfer. Main bath at 47.1%. Helium left in dewar: 6.2L. |
| 1720 | V9 closed with G2=349.8mbar. V12A opened all the way (done slowly keeping G1<200mbar.) |
| 1800 | G1=152.8mbar and G2=145.7mbar. Given current rate of decrease in G1, it will take roughly 5 hours to bring G1 down low enough to start circulating. Thus I shall leave and (hopefully) return to start circulation. Depends on my ability to safely get back to work. Lakeshore/MC=17.9K. |
| 1920 | Fill LN2 cold trap full. |
| 2343 | G1=98.9, G2=92.8. Since it took almost 6 hours to get down 50mbar, I imagine it will take at least 2+hours to get down to the recommended 80mbar. Thus I will just wait till morning. We can start circulating as we do our transfer and then starting flashing detectors and taking data immediately. |