| 1003 | Mixture still appears to be flowing (i.e. cryocleaning). Currently G1=546.1mbar and G2=504.2mbar. |
| 1046 | Filled LN2 coldtrap. |
| 1155 | Begin pumping mixture to dump. |
| 1151 | Isolate dump by closing all valves on IGH, G1=0.4mbar, G2=694.6mbar. |
| 1221 | Begin helium transfer. Main bath empty (at -1%). |
| 1315 | Ran out of He gas, borrowed some from Zudov's lab. Continued LHe transfer (-1%). |
| 1356 | End helium transfer. Main bath 92.1%. |
| 1431 | Started 1K Pot. Returned He gas to Zudov's lab. |
| 1500 | Begin condensing mixture. |
| 1820 | Current status: Lakeshore/MC=47.3K, G1=226.9mbar G2=559.0mbar. The mixture has been condensing with G1>200mbar for a few hours. Although it is now decreasing, it is doing so very slowly, most likely due to the high MC temp that is also currently decreasing. For now it will be left to its own devices for a few hours. |
| 2059 | Current status: Lakeshore/MC=44.0K, G1=226.0mbar G2=559.5mbar. The mixture has been condensing with G1>200mbar for a many hours, as seen in the last entry. As mentioned before, I believe this is due to the high MC temp. Once this drops, I am hopeful that it will condense. Then again, I am somewhat new at this. |
| 2107 | Filled LN2 coldtrap full. |