2013/06/25
| 1006 | Filled LN2 cold trap full. |
|---|---|
| 1042 | Begin LHe transfer (-1.0%). |
| 1200 | Electrical check showed all OK. |
| Cooldown checkout sheet (rev C SQUET) | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Detector: | G101b DIB1 | BH: | Stripline: | 27 | |||||||
| SC: | 300 K | 80K | LHe | ||||||||
| Face: | 2 | Date | 2012-06-17 | Date | 2012-06-23 | Date | 2013-06-25 | ||||
| SQUET: | ZC23 | Who | AC, HR, CP | Who | HC | Who | AK | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.8k | OK | 1 - 2 k | 1.9k | OK | 80 - 140 | 97.5 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.3k | OK | 1.7k | OK | 80 - 140 | 103.1 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.6k | OK | 1.8k | OK | 80 - 140 | 94.8 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 5.2k | OK | 2.1k | OK | 80 - 140 | 94.2 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 29k | OK | 9 - 13 k | 15k | OK | 1.0 - 1.3 k | 1.3k | OK |
| Feedback B/Ret | 2 | 18 | 25k | OK | 9.1k | OK | 1.0 - 1.3 k | 1.2k | OK | ||
| Feedback C/Ret | 41 | 25 | open | OK | open | OK | 1.0 - 1.3 k | open | OK | ||
| Feedback D/Ret | 22 | 6 | 35k | OK | 17k | OK | 1.0 - 1.3 k | 1.2k | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 98, 87, 82, 85, open | OK | 30 - 40 | 110,61,47,49,open | OK | 20 - 30 | 38,39,25,27,open | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 81, 80, 83, 80 | OK | 41,40,55,40 | OK | 20 - 30 | 24, 23, 39, 23 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 81, 81, 72, 80 | OK | 51,44,36,44 | OK | 20 - 30 | 31, 23, 15, 22 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 85, open, 82, 83 | OK | 48,open,45,46 | OK | 20 - 30 | 25, open, 23, 23 | OK | ||
| QET C/QET A Ret | 7 | 1, 13, 14, 17, 32 | 110,93,88,91,89 | ||||||||
| QET C/QET D Ret | 7 | 36, 40, 43, 44 | 90,87,86,87 | ||||||||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.06k | OK | 1.0 - 1.1 k | 1.04k | OK | 1.0 - 1.1 k | 1.23k | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | open(A, B, C), 26(D) | OK | < 1 nS | OK | OK | <1 nS | OK | OK | |
| Qo bias/QET Ret | 11 | open(A, B, C), 26(D) | OK | OK | OK | <1 nS | OK | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23,23 | OK | 6 - 8 nS | 19,20 | OK | <1 nS | 19, 19 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23,23 | OK | 20,20 | OK | <1 nS | 20, 20 | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | OK | <0.1 nS | – | OK | |
| Qo drain | 12 | chassis | – | OK | OK | <0.1 nS | – | OK | |||
| Qi source | 31 | chassis | – | OK | OK | <0.1 nS | – | OK | |||
| Qo source | 27 | chassis | – | OK | OK | <0.1 nS | – | OK | |||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | open | open | 0.9 - 1.0 V | 1.09 | OK | — | OK | |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.135 | – | 1.2 - 1.5 V | 1.415 | – | — | – | |
| LED 2 | 26 | chassis | open | – | open | – | — | – | |||
| Cooldown checkout sheet (rev C SQUET) | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Detector: | G101b DIB2 | BH: | Stripline: | 19 | |||||||
| SC: | ZB1-1-9 | 300 K | LN | LHe | |||||||
| Face: | 5 | Date | 2013-06-17 | Date | 2013-06-23 | Date | 2013-06-25 | ||||
| SQUET: | ZC23 | Who | AC, HR, CP | Who | HC | Who | AK | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4k | OK | 1 - 2 k | 1.6k | OK | 80 - 140 | 104 | OK |
| SQUID bias B/Ret | 19 | 34 | 5.0k | OK | 2.0k | OK | 80 - 140 | 94.3 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.6k | OK | 1.7k | OK | 80 - 140 | 101.9 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 4.3k | OK | 1.6k | OK | 80 - 140 | 92.5 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 18k | OK | 9 - 13 k | 7.3k | OK | 1.0 - 1.3 k | 1.3 k | OK |
| Feedback B/Ret | 2 | 18 | 20k | OK | 8.2k | OK | 1.0 - 1.3 k | 1.3k | OK | ||
| Feedback C/Ret | 41 | 25 | 19k | OK | 7.6k | OK | 1.0 - 1.3 k | 1.3k | OK | ||
| Feedback D/Ret | 22 | 6 | 18k | OK | 7.0k | OK | 1.0 - 1.3 k | 1.3k | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 102, 90, 85, 87,86 | OK | 30 - 40 | 87,64,49,52,50 | OK | 20 - 30 | 76, 45, 32, 34, 33 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 87, 86, 91, 87 | OK | 47,46,61,47 | OK | 20 - 30 | 29, 28, 46, 29 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 86, 86, 74, 85 | OK | 56,47,36,46 | OK | 20 - 30 | 38, 30, 19, 29 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 86, 84, 83, 84 | OK | 48,46,45,46 | OK | 20 - 30 | 31, 29, 28, 29 | OK | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.07k | OK | 1.0 - 1.1 k | 1.05k | OK | 1.0 - 1.1 k | 1.24k | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | open (A,B,C); 26 (D) | OK | < 1 nS | open | OK | <1 nS | open | OK | |
| Qo bias/QET Ret | 11 | open (A,B,C); 25 (D) | OK | open | OK | <1 nS | open | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23,23 | OK | 6 - 8 nS | 21,21 | OK | <1 nS | 18, 18 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23,23 | OK | 21,21 | OK | <1 nS | 19, 19 | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.727 | OK | 0.9 - 1.0 V | 1.092 | OK | — | 1.167 | OK |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.14 | – | 1.2 - 1.5 V | 1.419 | – | — | – | |
| LED 2 | 26 | chassis | 1.13 | – | 1.409 | – | — | – | |||
| 1230 | Finish LHe transer (91.9 %). |
|---|---|
| 1455 | MC thermometer: 2.1K. Start circulation procedure. |
| 1812 | DCRC32 was attached to Dib2 (stripline 19) and DCRC33 was attached to Dib1 (stripline 27). |
| 1922 | Filled LN2 cold trap. |
| 1925 | DCRC33 would not connect for an unknown reason and was thus removed. DCRC32 was then attached to Dib1 (27) and DCRC31 was attached to Dib2 (19). They appear to be working properly but with some noise. Mixing chamber temp at 68.3mK. LED 1 used for Dib1 and LED 1 and 2 used for Dib2. |