2013/06/06
| 1200 | Completed tower assembly and electrical check. Electrical check produced mostly expected results except that conductance values for LED 2 on Dib 2 continue to increase, when it should be below 0.1 nS. |
|---|
| + Lead | - Lead | Dib 1 value | Dib 2 value |
|---|---|---|---|
| 1 | 2 | open | open |
| 2 | 4 | open | open |
| 3 | 4 | open | open |
| 4 | 6 | open | open |
| 5 | 6 | open | open |
| 6 | 8 | open | open |
| 7 | 8 | 333 Ohms | 342 Ohms |
| 2 | 8 | open | open |
| Chassis | 9 (RF) | 769 kOhms | 786 kOhms |
| Chassis | 9 (VF) | 1.103 V | 1.106 V |
| 9 | Chassis | < 1 nS | < 1 nS |
| Chassis | 10 (RF) | Open | 780 kOhms |
| Chassis | 10 (VF) | Open | 1.109 V |
| 10 | Chassis) | < 1 nS | < 1 nS |
| 7 | 11 | 25 nS | 26 nS |
| 12 | 11 | < 1 nS | < 1 nS |
| 12 | 13 | 25.5 nS | 26 nS |
| 7 | 14 | 26 nS | 26 nS |
| 15 | 14 | < 1 nS | < 1 nS |
| 15 | 16 | 25.5 nS | 26 nS |