start › cdms › k100 › run_diary › 20130330

2013/03/30

1200 Close V1, V2. Open vent speedivalve. G2 ~ 7.4 mbar and decreasing. Turn off He-3 pump. Internal leak rate of leak checker: 1.03E-10 mbar*l/s.
1212 Isolated He4 pump, opened leak checker to fridge lines. Leak rate rose to 1.8E-4 mbar*l/s and then slowly dropped. G2 ~ 4.4 mbar.
1220 Close V13A and B. Saw G2 rise.
1235 Installed styrofoam cup phone, aka “Styrophone”. Functional and helpful! G2 at 23 mbar and leak rate is 1.2E-8 mbar*l/s.
Begin leak checking by opening V13A and V13B. Leak rate rose to 1.4E-4 mbar*l/s.
1242 G2 at 11.2 mbar.
First leak test conducted with syringe. The highest two flanges + the speedivalve being sprayed seemed to result in a rapid increase of the leak rate to 1.6E-4 mbar*l/s. Stopping the spray test, leak rate lowered to 1.3E-4 mbar*l/s. It seems that at this rate the leak rate background started to decrease more steadily.
1245 Leak rate has decreased to 1.04E-4 mbar*l/s. G2 at 3.8 mbar. The first round of checking all flanges above pump (and accessible by leak detector) produced no gross leak. Re-testing the top flanges/ valve did not reproduce the initial higher leak rate.
Background leak rate is 5.53E-5 mbar*l/s.
5-10 minutes later, conducted liberal spray test over all flanges, starting with a leak rate of 4.2E-5 mbar*l/s (and dropping).
1315 Leak rate 3.0E-5 mbar*l/s and decreasing. No sign of gross leak. A small increase was detected, but re-tracing steps did not reproduce it.
1324 Background leak rate: 1.97E-5 mbar l/s. G2 ~ 1.7 mbar.
1330 Switched spray operators. Background leak rate: 1.5E-5 mbar*l/s.
1345 Detected a large leak in valves right above pump. Maxed out at 4.6E-5 mbar*l/s. Should use the syringe again.
1410 Background leak rate: 1.76E-5 mbar*l/s, G2 ~ 2.3 mbar.
1420 Leak rate rose from 8.8E-6 mbar*l/s to 1.3E-5 mbar*l/s. Significant leak detected when spraying around the pump.
1445 Covered the end of the pump where the oil case panel is with plastic (taped. down). Also covered suspect flanges above pump with plastic. No picture, unfortunately.
Background leak rate: 9.8E-6 mbar*l/s. Rate increased to 2.4E-5 mbar*l/s due to spraying (without syringe) of the bottom of the oil case front panel.
1457 Leak rate maxed at 3.95E-5 mbar*l/s.
1510 Begin spraying on oil plugs on pump with syringe. Background rate: 1.12E-5 mbar*l/s.
1555 Slight increases in leak rate were detected several times ( ~ 0.1E-6 mbar*l/s) and then spraying would be stopped to try to determine location of leak. However, trying to localize the spot around the plugs proved to be very difficult.
1622 Covered back of the pump and over the motor with plastic. Placed the clear hose without syringe into back of pump above motor. The background leak rate started at 5.22E-6 mbar*l/s.
1625 A sharp increase in the leak rate occurred - not levelling off.
Turn off gas supply; G2 rose slightly ( ~ 3 mbar).
Leak rate maxed at 3.75E-4 mbar*l/s.
1700 Now using syringe to re-produce leak. Leaving syringe in above motor at back of pump. Current leak rate at the start of this test: 1.5E-4 mbar*l/s.
1706 Leak rate starting to increase.
1711 Stop gas supply. Leak rate at this point: 1.88E-4 mbar*l/s. This was high/different enough to indicate that a leak is definitely there.
Leak rate maxed at 2.24E-4 mbar*l/s.
1812 Isolate leak checker. Background seemed stable at 5.5E-5 mbar*l/s. Open He-4 pump to fridge lines again.