| 0947 | Mixing chamber temperature stable at 57.0 mK. G1 and G2 are 77 and 78 mbar, respectively. They've been slowly increasing overnight. However, I'm not alarmed anything will happen to them just now. |
| 0952 | Filled LN2 cold trap. |
| 1120 | Fed Cricket 4 pellets. |
| 1225 | Remove DCRC33 and attach DCRC32 to Dib2 (stripline 19). Begin flashing detector for 20 minutes, then cool down for 1 hour. Take one 30 min series at - 4 V for Dib 2. Trigger for Qi channel. |
| | Been having trouble seeing raw pulses. Need to check processing. |
| 1600 | Open V11B and V8 to pull out any mixture left in cold trap 2. |
| | Scroll pump has also been connected to vent valve, but vent valve is shut and so is V7. |
| 1603 | Shut V11B and V8. |
| | Decided that perhaps we should switch to circulating through cold trap 2 since we've been circulating through cold trap 1 for a day now. |
| 1610 | P1 = 0.07 mbar. |
| 1611 | Open V13B. This lets mixture into cold trap 2 now. |
| 1615 | Open V12B. Now circulating through both cold traps. |
| 1618 | Close V13A, 12A. |
| 1620 | P1 ~ 0.065 - 0.07 mbar. Open V11A, V8. Saw pressure rise in P1 to 30 mbar. |
| 1705 | P1 ~ 0.09 mbar. Closed V11A. P1 increased to 8 mbar. Re-opened V11A to keep pumping mixture to circulation line. |
| 1750 | P1 has stabilized at 0.15 mbar for half an hour. It seems like all the mixture is out, though the pressure in P1 is higher. |
| 1755 | Pull out cold trap 1. Turn on scroll pump and evacuate up to vent valve. |
| | Oh my goodness… there is an awful smell. Apparently that's coming from the cold trap as it's being warmed up, but this has never been noticeable before. |
| | Open vent valve to evacuate. Open 7 and 11A to start pumping on cold trap. |
| 1820 | Fed Cricket 4 pellets. |
| 1942 | In the midst of 2nd cold trap cleaning. Mixing chamber temperature is 60 mK. P1 ~ 0.085 mbar. G1 ~ 84.4 mbar, G2 ~ 77.6 mbar. G1 and G2 seem relatively stable. |
| 2015 | Finished 3 rounds of cleaning cold trap 1. |
| 2100 | Closed off cold trap 1 and returned it to LN2 dewar. |
| 2125 | Begin LHe transfer (35.9%). |
| 2155 | End LHe transfer (93.1%). |