2013/03/10
| 0010 | Mixing chamber temperature has been rising for the last 10 minutes at a decent rate. Mixing chamber temperature is 7.1 K, up from 6.5 K. Hook up helium gas supply to liquid helium dewar to begin fast transfer. |
|---|---|
| 0020 | Resume mixture cryocleaning through nitrogen cold trap 2. |
| 0055 | Cease helium transfer. Helium liquid level 91.1%. |
| 1430 | Cryocleaning stagnated overnight. Open valve V6 on IGH to 14.3%. |
| Fill N2 cold trap full. | |
| Mixing chamber temperature 19.7 K. It's rather high, but this simply indicates that there is no gross leak. Isolate IVC and start 1 K pot: temperature 1.7 K. | |
| 1440 | Very quick electrical check for DIB 1 shows that QET D Returns are still connected as are Qi & Qo Feedbacks, Drains & Sources. Qi & Qo biases still read open when checked to QET D/A Returns, but this is expected and probably does not indicate a problem. |
| 1505 | Found mixing chamber heater had been left on, hence the higher than normal mixing chamber temperature. Switched mixing chamber heater off: now I'm unsure of whether we have a leak. If we didn't see one at 77 K, and considering that this is an ionisation-only run and we can live with a higher base temperature, chances are we are fine. |
| 1715 | Mixing chamber temperature 18.0 K and falling. Temperature of 1 K pot 1.8 K and holding steady. Interrupt cryocleaning and pump mixture into dump in preparation for condensation. |
| Fed Cricket 4 pellets. | |
| 1810 | Dump pressure 725 mbar. Mxing chamber temperature 17.4 K. Start manually condensing mixture through nitrogen cold trap 1. |
| 1815 | Fill N2 cold trap full. |
| 1830 | Poly-lead shield erected around cryostat. |
| 2030 | Mixing chamber temperature 11.8 K and falling steadily. Leave 376 mbar of mixture in dump (ie. around 349 mbar in fridge). Continue manual condensation procedure; start opening valve V12A on IGH. |
| 2245 | Mixing chamber temperature 6.3 K and falling steadily. Pressures in P1 & P2 at 120 mbar & 115 mbar respectively. Open valve V12A to 100% to end manual condensation procedure. Wait for mixing chamber to cool further. |
| 2355 | Conduct electrical check as mixing chamber cools from 4.2 K to 3.6 K. Only change from that at 77 K is the recovery of Feedback A to Return. Otherwise, as far as we can tell, the important channels are still in place. |
| Detector: | G101b DIB1 | BH: | Stripline: | 21 | |||||||
| SC: | ZB1-1-6 | 300 K | 80K | LHe | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 2 | Date | 2012-03-07 | Date | 2012-03-09 | Date | 2013-03-10 | ||||
| SQUET: | ZC70 | Who | HC | Who | HC | Who | HC | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | open | OK | 1 - 2 k | open | OK | 80 - 140 | open | OK |
| SQUID bias B/Ret | 19 | 34 | 4.4k | OK | open | OK | 80 - 140 | open | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.7k | OK | 1.8k | OK | 80 - 140 | 99 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 5.2k | OK | open | OK | 80 - 140 | open | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 29k | OK | 9 - 13 k | 2.2M | OK | 1.0 - 1.3 k | 1.5k | OK |
| Feedback B/Ret | 2 | 18 | 25k | OK | open | OK | 1.0 - 1.3 k | open | OK | ||
| Feedback C/Ret | 41 | 25 | 28k | OK | open | OK | 1.0 - 1.3 k | open | OK | ||
| Feedback D/Ret | 22 | 6 | 35k | OK | 17k | OK | 1.0 - 1.3 k | 1.2k | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 166,151,145,149,147 | OK | 30 - 40 | open | OK | 20 - 30 | open | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 84,83,97,83 | OK | 42,41,56,42 | OK | 20 - 30 | 28,27,43,27 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 157,85,76,85 | OK | open,46,37,46 | OK | 20 - 30 | open,27,18,27 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 93,90,89,90 | OK | open | OK | 20 - 30 | open | OK | ||
| QET C/QET A Ret | 7 | 1, 13, 14, 17, 32 | 110,93,88,91,89 | open,52,47,51,48 | open,32,27,30,29 | ||||||
| QET C/QET D Ret | 7 | 36, 40, 43, 44 | 90,87,86,87 | open,open,46,47 | open,open,26,27 | ||||||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.07k | OK | 1.0 - 1.1 k | 1.04k | OK | 1.0 - 1.1 k | 1.25k | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | open (B); 26 (A,C,D) | OK | < 1 nS | OK | OK | <1 nS | OK | OK | |
| Qo bias/QET Ret | 11 | open (B); 26 (A,C,D) | OK | OK | OK | <1 nS | OK | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23,23 | OK | 6 - 8 nS | 20,20 | OK | <1 nS | 18,18 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23,23 | OK | 21,21 | OK | <1 nS | 19,19 | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.715 | OK | 0.9 - 1.0 V | 1.094 | OK | — | OK | |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.135 | – | 1.2 - 1.5 V | open | – | — | – | |
| LED 2 | 26 | chassis | open | – | open | – | — | – | |||
| Isolation Checks | |||||||||||
| QET C/Feedback D | 18 | 17 | <0.1 nS | OK | – | OK | OK | ||||
| QET C/SQUID bias D | 18 | 30 | OK | – | OK | OK | |||||
| QET D/Feedback D | 29 | 17 | OK | – | OK | OK | |||||
| QET D/SQUID bias D | 29 | 30 | OK | – | OK | OK | |||||
| QET A Ret/QET B Ret | 1, 13, 14, 17, 32 | 29, 33, 47, 48 | OK | – | OK | OK | |||||
| QET A Ret/Qi feedback | 1, 13, 14, 17, 32 | 38 | OK | – | OK | OK | |||||
| QET A Ret/Qi drain | 1, 13, 14, 17, 32 | 25 | OK | – | OK | OK | |||||
| QET D Ret/QET B Ret | 36, 40, 43, 44 | 29, 33, 47, 48 | OK | – | OK | OK | |||||
| QET D Ret/Qi feedback | 36, 40, 43, 44 | 38 | OK | – | OK | OK | |||||
| QET D Ret/Qi drain | 36, 40, 43, 44 | 25 | OK | – | OK | OK | |||||
| Detector: | G101b DIB2 | BH: | Stripline: | 19 | |||||||
| SC: | ZB1-1-9 | 300 K | LN | LHe | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 5 | Date | 2013-03-07 | Date | 2013-03-09 | Date | 2013-03-10 | ||||
| SQUET: | ZC23 | Who | HC | Who | HC | Who | hc | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4k | OK | 1 - 2 k | 1.6k | OK | 80 - 140 | 107 | OK |
| SQUID bias B/Ret | 19 | 34 | 5.0k | OK | 2.0k | OK | 80 - 140 | 97 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.6k | OK | 1.7k | OK | 80 - 140 | 105 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 4.3k | OK | 1.6k | OK | 80 - 140 | 95 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 18k | OK | 9 - 13 k | 7.3k | OK | 1.0 - 1.3 k | 1.3k | OK |
| Feedback B/Ret | 2 | 18 | 20k | OK | 8.3k | OK | 1.0 - 1.3 k | 1.3k | OK | ||
| Feedback C/Ret | 41 | 25 | 19k | OK | 7.6k | OK | 1.0 - 1.3 k | 1.3k | OK | ||
| Feedback D/Ret | 22 | 6 | 18k | OK | 7.1k | OK | 1.0 - 1.3 k | 1.3k | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 90,91,85,89,86 | OK | 30 - 40 | 51,50,45,48,46 | OK | 20 - 30 | 38,37,32,35,34 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 87,86,100,86 | OK | 46,45,60,46 | OK | 20 - 30 | 33,32,48,32 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 85,85,75,85 | OK | 45,45,34,44 | OK | 20 - 30 | 32,33,22,32 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 86,84,84,84 | OK | 47,44,44,44 | OK | 20 - 30 | 34,32,31,32 | OK | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.07k | OK | 1.0 - 1.1 k | 1.04k | OK | 1.0 - 1.1 k | 1.27k | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | open (A,B,C); 26 (D) | OK | < 1 nS | OK | OK | <1 nS | OK | OK | |
| Qo bias/QET Ret | 11 | open (A,B,C); 25 (D) | OK | OK | OK | <1 nS | OK | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23,23 | OK | 6 - 8 nS | 20,20 | OK | <1 nS | 18,18 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23,23 | OK | 20,20 | OK | <1 nS | 19,19 | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.725 | OK | 0.9 - 1.0 V | 1.092 | OK | — | OK | |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.137 | – | 1.2 - 1.5 V | 1.414 | – | — | – | |
| LED 2 | 26 | chassis | 1.13 | – | 1.404 | – | — | – | |||