2013/02/24
| 0450 | Mixing chamber temperature 10 K. Pump helium exchange gas out of IVC. |
|---|---|
| 0740 | Mixing chamber temperature 12.4 K. Start filling main bath with liquid helium. |
| 0830 | Cease liquid helium transfer. Helium liquid level 92.1%. |
| 0945 | Fill N2 cold trap full. Restart cryocleaning mixture through nitrogen cold trap 2. |
| Fed Cricket 3 pellets. | |
| 1825 | Mixing chamber temperature 11.2 K and decreasing very slowly. It was around 14 K at end of helium transfer. This doesn't necessarily indicate that there is a leak as the 1 K pot's needle valve is known to leak, and therefore it's temperature is likely to be lower. Temperature of 1 K pot is greater than 7 K. |
| 1830 | Mixture has stagnated. Fill N2 cold trap full. Open valve V6 on IGH to 13.9% and resume cryocleaning. |
| 1955 | Pump mixture into dump in preparation for condensation. |
| 2000 | Isolate IVC. Start 1 K pot; temperature 1.83 K. |
| 2050 | Start condensing mixture manually through nitrogen cold trap 1. |
| 2115 | Cease manual condensation procedure. Pressure in dump before condensation 720 mbar; after condensation 369 mbar: around 351 mbar of mixture fed into fridge. Open valve 12A on IGH to 100%. Mixing chamber temperature 9.5 K and falling at what appears to be an increasing pace. |
| 2355 | Conduct electrical check (mixing chamber temperature dropped from 4.1 K to 3.4 K during check). SQUID Bias to Return for A, B & D on DIB 3 still open; Feedback A & B to Return on DIB 3 still open; QET A & D to Return on DIB 3 still open: fortunately, channel C on DIB 3 looks fine so we can still read all the phonon channels on Side 1. LED 2 on DIB 3 leaks in reverse direction, but this is a known problem: lower conductivity of 13 nS than last run (24 nS). |
| Detector: | G106a DIB1 | BH: | Stripline: | 19 | ||||||||
| SC: | ZB1-4-6 | 300 K | 80K | LHe | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 2 | Date | 2013-02-20 | Date | 2013-02-23 | Date | 2013-02-24 | |||||
| SQUET: | ZC23 | Who | AK & SM | Who | RR | Who | HC | |||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis | |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | ||||||||||
| Nominal: | Nominal: | Nominal: | ||||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | ||||||||||
| Phonon | ||||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4k | OK | 1 - 2 k | 1.7k | OK | 80 - 140 | 107 | OK | |
| SQUID bias B/Ret | 19 | 34 | 5.0k | OK | 2.0k | OK | 80 - 140 | 96 | OK | |||
| SQUID bias C/Ret | 8 | 24 | 4.6k | OK | 1.8k | OK | 80 - 140 | 105 | OK | |||
| SQUID bias D/Ret | 23 | 39 | 4.3k | OK | 1.6k | OK | 80 - 140 | 95 | OK | |||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 19k | OK | 9 - 13 k | 7.4k | OK | 1.0 - 1.3 k | 1.27k | OK | |
| Feedback B/Ret | 2 | 18 | 20k | OK | 8.4k | OK | 1.0 - 1.3 k | 1.27k | OK | |||
| Feedback C/Ret | 41 | 25 | 19k | OK | 7.7k | OK | 1.0 - 1.3 k | 1.26k | OK | |||
| Feedback D/Ret | 22 | 6 | 18k | OK | 7.2k | OK | 1.0 - 1.3 k | 1.25k | OK | |||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 111,97,89,92,90 | OK | 30 - 40 | 52, 51, 46, 50, 48 | OK | 20 - 30 | 37,36,31,35,34 | OK | |
| QET B/Ret | 35 | 29, 33, 47, 48 | 90,90,98,90 | OK | 48, 47, 69, 47 | OK | 20 - 30 | 33,32,51,32 | OK | |||
| QET C/Ret | 7 | 5, 9, 10, 28 | 86,85,75,85 | OK | 47, 47, 36,46 | OK | 20 - 30 | 32,33,22,32 | OK | |||
| QET D/Ret | 38 | 36, 40, 43, 44 | 100,86,85,86 | OK | 48, 46, 46 ,46 | OK | 20 - 30 | 34,32,31,31 | OK | |||
| Charge | ||||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.07k | OK | 1.0 - 1.1 k | 1.045k | OK | 1.0 - 1.1 k | 1.26k | OK | |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 25 | OK | < 1 nS | OK | OK | <1 nS | OK | OK | ||
| Qo bias/QET Ret | 11 | 25 | OK | OK | OK | <1 nS | OK | OK | ||||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23,23 | OK | 6 - 8 nS | 20,20 | OK | <1 nS | 18,18 | OK | |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23,23 | OK | 20,20 | OK | <1 nS | 19,19 | OK | |||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK | |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.723 | OK | 0.9 - 1.0 V | 1.09 | OK | — | – | OK | |
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.128 | – | 1.2 - 1.5 V | 1.4 | – | — | – | – | |
| LED 2 | chassis | 26 | 1.117 | – | 1.392 | – | — | – | – | |||
| Detector: | G106a DIB2 | BH: | Stripline: | 20 | |||||||
| SC: | ZB1-4-4 | 300 K | LN | LHe | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 5 | Date | 2013-02-20 | Date | 2013-02-23 | Date | 2013-02-24 | ||||
| SQUET: | ZC70 | Who | AK & SM | Who | RR | Who | HC | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.8k | OK | 1 - 2 k | 2.0k | OK | 80 - 140 | 99 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.4k | OK | 1.7k | OK | 80 - 140 | 105 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.6k | OK | 1.8k | OK | 80 - 140 | 96 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 5.2k | OK | 2.1k | OK | 80 - 140 | 96 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 29k | OK | 9 - 13 k | 15k | OK | 1.0 - 1.3 k | 1.25k | OK |
| Feedback B/Ret | 2 | 18 | 25k | OK | 9.3k | OK | 1.0 - 1.3 k | 1.25k | OK | ||
| Feedback C/Ret | 41 | 25 | 28k | OK | 14k | OK | 1.0 - 1.3 k | 1.26k | OK | ||
| Feedback D/Ret | 22 | 6 | 34k | OK | 17k | OK | 1.0 - 1.3 k | 1.24k | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 110,96,83,86,.85 | OK | 30 - 40 | 50, 49, 44, 47, 46 | OK | 20 - 30 | 29,28,23,26,25 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 81,80,94,80 | OK | 40, 39, 60, 39 | OK | 20 - 30 | 25,24,43,25 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 70,77,70,77 | OK | 43, 43, 35, 42 | OK | 20 - 30 | 24,24,16,23 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 100,82,83,82 | OK | 47, 45, 44, 45 | OK | 20 - 30 | 26,25,24,24 | OK | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.06k | OK | 1.0 - 1.1 k | 1.042k | OK | 1.0 - 1.1 k | 1.24k | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 26 | OK | < 1 nS | OK | OK | <1 nS | OK | OK | |
| Qo bias/QET Ret | 11 | 26 | OK | OK | OK | <1 nS | OK | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 24,24 | OK | 6 - 8 nS | 21,21 | OK | <1 nS | 19,19 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 24,24 | OK | 21,21 | OK | <1 nS | 19,19 | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | open | OK | 0.9 - 1.0 V | 1.084 | OK | — | – | OK |
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.12 | – | 1.2 - 1.5 V | 1.397 | – | — | – | – |
| LED 2 | chassis | 26 | 1.11 | – | 1.39 | – | — | – | – | ||
| Detector: | G106a DIB3 | BH: | Stripline: | 21 | |||||||
| SC: | ZB1-4-8 | 300 K | LN | LHe | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 6 | Date | 2013-02-20 | Date | 2013-02-23 | Date | 2013-02-24 | ||||
| SQUET: | ZC13 | Who | AK & SM | Who | RR | Who | HC | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.6k | OK | 1 - 2 k | open | OK | 80 - 140 | open | OK |
| SQUID bias B/Ret | 19 | 34 | 4.7k | OK | open | OK | 80 - 140 | open | OK | ||
| SQUID bias C/Ret | 8 | 24 | 5.3k | OK | 2.1k | OK | 80 - 140 | 118 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 4.9k | OK | open | OK | 80 - 140 | open | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 19k | OK | 9 - 13 k | open | OK | 1.0 - 1.3 k | open | OK |
| Feedback B/Ret | 2 | 18 | 20k | OK | open | OK | 1.0 - 1.3 k | open | OK | ||
| Feedback C/Ret | 41 | 25 | 30k | OK | 12k | OK | 1.0 - 1.3 k | 1.26k | OK | ||
| Feedback D/Ret | 22 | 6 | 21k | OK | 8.6k | OK | 1.0 - 1.3 k | 1.25k | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 84, 84, 79, 83, 81 | OK | 30 - 40 | open | OK | 20 - 30 | open | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 79, 79, 95, 79 | OK | 43, 42, 61, 42 | OK | 20 - 30 | 27,26,55,27 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 82,82,73,82 | OK | 42, 42, 33, 42 | OK | 20 - 30 | 27,27,18,27 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 83, 80, 79, 80 | OK | open | OK | 20 - 30 | open | OK | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.068k | OK | 1.0 - 1.1 k | 1.05k | OK | 1.0 - 1.1 k | 1.24k | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | <1 | OK | < 1 nS | open | OK | <1 nS | open | OK | |
| Qo bias/QET Ret | 11 | <1 | OK | open | OK | <1 nS | open | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23,23 | OK | 6 - 8 nS | 19, 19 | OK | <1 nS | 19,19 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23,23 | OK | 20,20 | OK | <1 nS | 19,19 | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.713 | OK | 0.9 - 1.0 V | 1.08 | OK | — | – | OK |
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.117 | 10 nS | 1.2 - 1.5 V | 1.396 | OK | — | – | – |
| LED 2 | chassis | 26 | 1.108 | open | 1.386 | 23 nS | — | – | 13 nS | ||