2013/02/17

0029 Begin taking data. First flash for 20 min then cool for 90 min. Take data with Dibs 1/2 biased at +/- 4 V. Trigger settings can be found in Tables 2-4 on yesterday's run diary. 15 min series, 6 seconds of flashing in between, and 20 minutes of cool down. 15 data series to be taken. Am-241 only.
1050 Filled LN2 cold trap.
Fed Cricket 4 pellets.
1120 Begin LHe transfer. (23.5%)
1126 Flash LEDs on Dibs 1/2 for 30 seconds.
1200 End LHe transfer. (83.1%)
1314 Data quality of this morning's data is not good. For some reason Qi and Qo on Dib 1 have no pulses showing. Need to figure out why…
1324 Reading from DCRC33 displays: 0B00. I think this is saying the FET heater was on, and I'm really unsure as to why that would be the case. Perhaps this had something to do with the charge signal disappearing.
1508 DAQ will not trigger past the initial first 500 events.
1528 Noticed that DCRC31 was connected to 2 ports and quit one. That didn't fix the triggering. When I reset the power to the boards, I found that Board 116 was still connected to the readout vi. This could have all been due to that.
1557 Excellent. The DAQ is triggering past the first 500 events again. Dib 2/3 still looking noisy. Still need to see why there were no signals Qi/Qo for Dib 1 in the data and if this happens again.
1605 Checked trigger channels. They seem acceptable. Keeping the same settings as this morning, did a quick data run. Qi and Qo definitely are showing pulses on Dib 1 and Dib 2. Data rate ~ 7 Hz.
1613 Take test data run. Flash for 6 seconds, cool for 20 min. One 15 min series with previous trigger settings.
1648 Checked on data taking. An error with board 32. When quitting the data acquisition process, board 32 stays connected to port 5000.
1650 Cleared port 5000 for board 32. Very quick manual flash. Manually started data taking with MultiDAQ vi. Charge pulses are visible. I'll let this run for 20 min.
Suspicious of the pre-flashing request made via the DAQDAQ2 vi. Maybe when it first flashed, it did not properly transition to data taking.
Note: Before data taking, 1) always check control/readout vi's to make sure they're off; 2) Type “sock” into putty session to make sure the boards are only connected to 5002 at the moment; 3) Type “rd 30” - this tells you the boards power status, whether it's zapping, flashing, etc. You're fine to take start taking data if it reads “0800” or “0A00”.
1852 Fortunately the pulses themselves look okay but the cold traps are dirty again. There's either a hard block or the mixture itself is really dirty. Warming up the fridge now.
Remove DCRC boards.
Pump mixture to dump.
Stop 1K pot.
Switch recovery to exhaust.
1910 Filled LN2 cold trap.
Turn on mixing chamber heater to 10 mW.
1956 Evacuated 1K pot.
2008 Fed Cricket too many pellets because I enjoyed watching him jump out of the water. He hit my finger trying to get the pellet! (4 < too many < 7 pellets)