start › cdms › k100 › run_diary › 20130206

2012/02/06

1027 Fed Cricket 4 pellets.
1033 Filled LN2 cold trap.
1050 Begin taking data. First flash detector for 6 seconds with both LEDs on Dibs 1/2. Cool down for 25 min. Then take data with Dib 1/2 biased at -/+ 3 V, 2.5 V, 2 V, 1.5 V, 1 V, and 0.5 V. At each bias there is a 30 min series, followed by 6 seconds of flashing and 20 minutes of cool down. End with 30 min of flashing, 30 min of cool down. Trigger on Qi/Qo on Dib 1 only at 45/40 ADC bins, respectively.
1756 Filled LN2 cold trap.
1820 Fed Cricket.
1823 Begin taking data. Dib 1/2 biased at -/+ 10 V and 12 V. At each bias there is a 30 min series, followed by 6 seconds of flashing and 20 minutes of cool down. End with 30 min of flashing, 30 min of cool down. Trigger on Qi/Qo on Dib 1 only at 45/40 ADC bins, respectively.