| 1025 | Mixing chamber temperature 50.6 mK. Start zapping SQUIDs starting with DIB 1. |
| 1110 | DCRC32 is behaving weirdly. Zapping the SQUIDs did not result in a rise in temperature, indicating that it had disconnected. A quick check indicates that it must have turned itself off and on again. We waste far too much time on this board. |
| 1215 | Finished zapping SQUIDs. Start flashing detector for about an hour: LEDs 1 & 2 on DIBs 1 & 2 only; usual settings (-0.6 mA current, 100 μs flash on time, 5 ms flash off time). |
| 1315 | Stop flashing detector. |
| 1720 | DCRC33's 15 V power supply stopped after detector was flashed. I'm beginning to suspect that the switch is the cause of the problem. |
| 1750 | Brought out Ba-133 source. No significant improvement in data acquisition rate, which was rather surprising. This is further compounded when trying to take data: although trigger rate looks healthy under bias from DetectorReadout vi, trigger rate is poor when taking random triggers (approximately 5 Hz), and then starts triggering after taking 250 events, half-way to its usual 500. |
| | Unplug everything. Exit LabVIEW. Start again. |
| 1835 | Nothing seems to work. It's impossible to take data at the moment. I strongly suspect that the switch is at fault. We should check tomorrow by trying to locate another switch. Looks like there will be no yield plot tonight. |
| 1930 | Fed Cricket 4 pellets. |
| 1935 | Fill N2 cold trap full. |