| 0000 | Start flashing detector: LEDs 1 & 2 on DIBs 1 & 2 only; usual settings (-0.6 mA current, 100 μs flash on time, 5 ms flash off time). |
| 0020 | Fill N2 cold trap full. |
| 0852 | Stop flashing detector (temperature was at 168.8 mK). |
| 0855 | Filled LN2 cold trap. |
| 1005 | Fed Cricket 4 pellets. |
| 1341 | Zapped SQUIDs via DCRC32 twice. Started pulser to take Tc measurements and saw there was trapped flux. Stopped pulser and board disconnected and VIs froze. Quit Labview. |
| | Note that the same thing happened yesterday when stopping the pulser on DCRC32. |
| 1439 | After zapping the SQUIDs twice again there still seems to be trapped flux. |
| | Switched to DCRC33 to see if the pulser problem occurred there. Test sending a signal down QET A and then stopped it - worked fine. |
| 1545 | No amount of zapping seems to be correcting the trapped flux on dib 2. I started the pulser just to be sure and when I stopped it the program froze. |
| 1559 | Unplugged cable from both board and switch. Re-started LabVIEW again. Re-powered DCRC32 and now zapping SQUIDs. |
| 1711 | Fed Cricket 5 pellets. |
| 2257 | Finished taking Tc measurements for Ch.'s A, B, D on Dib 2 with DCRC32. Used 10 mVpp wave at 100 Hz. |
| | Restarted temp.txt and igh.txt files. |
| 2300 | Start flashing detector. |
| 2310 | Filled LN2 cold trap. |