| 1010 | Filled LN2 cold trap. |
| | Fed Cricket 4 pellets. |
| 1014 | Mixing chamber temperature: 52.5 mK. Success! However, when leaving yesterday for the night to let fridge cool, the resistance range wasn't switched high enough, so we can't see the rate of decrease of the fridge's temperature. It was stuck at 90 mK. |
| 1045 | Begin LHe transfer (4.1%). |
| 1135 | End LHe transfer (87.5%). |
| 1322 | Mixing chamber temperature: 52.6 mK. |
| | Plugged in DCRC33. |
| 1412 | Thought switch wasn't working. It wasn't actually connected to the computer, but by the time I realized that I was already using the external PoE. So, board is being powered by external PoE. |
| | Started clock. |
| 1423 | External PoE randomly disconnected. Temperature rose 0.7 mK in the meantime (and still creeping up). |
| | Connection seems unstable. Keeps connecting and disconnecting. |
| 1431 | Re-connected board. Waiting to see if temperature levels off. |
| 1448 | Temperature has increased over 2 mK now, still increasing. Connection is finally stable. I think it takes awhile to reach this point. |
| 1458 | Temperature is fluctuating, but not going much higher than 54 mK. |
| 1515 | Temperature is stabilizing around 53.8 mK. Connection is still good. |
| 1537 | Begin tuning board (54 mK). |
| 1548 | Zap SQUIDs twice. |
| 1608 | Turning on test signal doesn't work. Not sure why. Board is powered, clock is started, and channels seem to be responding (perhaps a bit fast, but still seem fine). |
| 1642 | Filled LN2 cold trap full. |
| 1700 | fed Cricket 4 pellets. |