| 1010 | Fed Cricket 3 pellets. |
| 1015 | Filled LN2 cold trap. |
| | P1 = 0.013 mbar; G1 ~ 1.1 mbar; G2 ~ 725.5 mbar. |
| | Still missing about 4-5 mbar of mixture, and G1 had risen slightly. Also unsure of why it took so long to get this last 10 mbar out. |
| | We're going to continue pumping out the lines during the LHe transfer just in case there is mixture left, though the pressures suggest there isn't anything left. |
| 1050 | Warmed up and calibrated leak checker. Evacuating main bath right now. |
| 1054 | Background leak rate: 2.2E-9 mbar l/s. |
| 1100 | Pumped and flushed main bath. No gross leak detected. |
| 1104 | Evacuated 1 K pot. |
| 1119 | Entered 0.054 mbar of He exchange gas into IVC. |
| 1129 | Begin slow He transfer (0%). |
| 1300 | Temperature starting to decrease (had increased to 100 K). |
| 1652 | Closed off dump. G2 ~ 725 mbar. |
| 1724 | Pump out He exchange gas at 10 K. |
| 1947 | Mixture cleaning re-commenced. |
| 2000 | Filled LN2 cold trap. |
| 2015 | Begin applying external gas pressure to the LHe dewar. |
| 2105 | End LHe transfer (78.3%). |