2012/12/20
| 0015 | Fed Cricket 3 pellets. |
|---|---|
| Check of helium background leak rate in hose yields 4.1E-8 mbar l/s. Continue evacuating hose with turbo pump. IVC isolated. | |
| 0025 | Fill N2 cold trap full. |
| 1120 | Fed Cricket 3 pellets. |
| 1350 | Fill N2 cold trap full. |
| 1410 | Check of helium background leak rate in hose yields 3.2E-8 mbar l/s. Leak rate drops when IVC is opened. Isolate IVC. Continue evacuating hose with turbo pump. |
| Pump mixture into dump in preparation for condensation. | |
| 1430 | Start 1 K pot, temperature 1.76 K. |
| 1435 | Stop pumping mixture into dump. Pressure in dump 731 mbar. Open gate valve and condenser line speedivalve. Stabilise 1 K pot temperature at 1.83 K. Mixing chamber temperature 7.5 K. Start condensing mixture manually through nitrogen cold 1. |
| 1450 | Close off dump. Pressure of mixture left in dump 380 mbar (ie. 350 mbar dump pressure fed into fridge). |
| 1455 | Fully open valve V12A on IGH. Wait for mixture to condense. Mixing chamber temperature 7.1 K and falling. |
| 1700 | Leak checked lines, valves and gaskets from valve isolating IVC to leak checker. Started with helium background leak rate of 5.2E-8 mbar l/s, and finished with 1.7E-7 mbar l/s. Increase in line with that expected from steadily rising background. No sudden jumps in the leak rate were witnessed. |
| 1725 | Commence helium transfer. Helium liquid level 0.1%. |
| 1815 | Cease helium transfer. Helium liquid level 90.2%. |
| 2000 | Conduct electrical check at 2.2 K. No problems witnessed apart from LED 2 on DIB 3 closed (conductivity 23 nS) in reverse direction: |
| Detector: | G106a DIB1 | BH: | Stripline: | 19 | ||||||||
| SC: | ZB1-4-6 | 300 K | 80K | LHe | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 2 | Date | 2012-12-16 | Date | 2012-12-18 | Date | 2012-12-20 | |||||
| SQUET: | ZC23 | Who | HC & RR | Who | HC | Who | HC | |||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis | |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | ||||||||||
| Nominal: | Nominal: | Nominal: | ||||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | ||||||||||
| Phonon | ||||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4k | OK | 1 - 2 k | 1.6k | OK | 80 - 140 | 107 | OK | |
| SQUID bias B/Ret | 19 | 34 | 5.0k | OK | 2.0k | OK | 80 - 140 | 96 | OK | |||
| SQUID bias C/Ret | 8 | 24 | 4.6k | OK | 1.7k | OK | 80 - 140 | 105 | OK | |||
| SQUID bias D/Ret | 23 | 39 | 4.3k | OK | 1.6k | OK | 80 - 140 | 95 | OK | |||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 19k | OK | 9 - 13 k | 7.3k | OK | 1.0 - 1.3 k | 1.27k | OK | |
| Feedback B/Ret | 2 | 18 | 20k | OK | 8.2k | OK | 1.0 - 1.3 k | 1.27k | OK | |||
| Feedback C/Ret | 41 | 25 | 19k | OK | 7.6k | OK | 1.0 - 1.3 k | 1.26k | OK | |||
| Feedback D/Ret | 22 | 6 | 18k | OK | 7.1k | OK | 1.0 - 1.3 k | 1.25k | OK | |||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 107,115,88,92,89 | OK | 30 - 40 | 61,59,48,52,49 | OK | 20 - 30 | 50,47,35,38,36 | OK | |
| QET B/Ret | 35 | 29, 33, 47, 48 | 91,90,105,91 | OK | 50,49,63,49 | OK | 20 - 30 | 35,35,49,35 | OK | |||
| QET C/Ret | 7 | 5, 9, 10, 28 | 88,86,75,85 | OK | 47,47,36,46 | OK | 20 - 30 | 34,33,22,32 | OK | |||
| QET D/Ret | 38 | 36, 40, 43, 44 | 94,86,86,86 | OK | 52,48,47,47 | OK | 20 - 30 | 39,33,33,33 | OK | |||
| Charge | ||||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.07k | OK | 1.0 - 1.1 k | 1.045k | OK | 1.0 - 1.1 k | 1.26k | OK | |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 25 | OK | < 1 nS | OK | OK | <1 nS | OK | OK | ||
| Qo bias/QET Ret | 11 | 25 | OK | OK | OK | <1 nS | OK | OK | ||||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23,23 | OK | 6 - 8 nS | 20,20 | OK | <1 nS | 18,18 | OK | |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23,23 | OK | 20,20 | OK | <1 nS | 19,19 | OK | |||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK | |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.723 | OK | 0.9 - 1.0 V | 1.094 | OK | — | – | OK | |
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.128 | – | 1.2 - 1.5 V | 1.403 | – | — | – | – | |
| LED 2 | chassis | 26 | 1.117 | – | 1.395 | – | — | – | – | |||
| Detector: | G106a DIB2 | BH: | Stripline: | 20 | |||||||
| SC: | ZB1-4-4 | 300 K | LN | LHe | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 5 | Date | 2012-12-16 | Date | 2012-12-18 | Date | 2012-12-20 | ||||
| SQUET: | ZC28 | Who | HC & RR | Who | HC | Who | HC | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.0k | OK | 1 - 2 k | 1.5k | OK | 80 - 140 | 113 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.3k | OK | 1.5k | OK | 80 - 140 | 91 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.9k | OK | 1.9k | OK | 80 - 140 | 102 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 4.3k | OK | 1.5k | OK | 80 - 140 | 88 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 24k | OK | 9 - 13 k | 8.4k | OK | 1.0 - 1.3 k | 1.25k | OK |
| Feedback B/Ret | 2 | 18 | 17k | OK | 7.2k | OK | 1.0 - 1.3 k | 1.26k | OK | ||
| Feedback C/Ret | 41 | 25 | 21k | OK | 8.6k | OK | 1.0 - 1.3 k | 1.42k | OK | ||
| Feedback D/Ret | 22 | 6 | 18k | OK | 7.1k | OK | 1.0 - 1.3 k | 1.25k | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 94,94,77,80,.77 | OK | 30 - 40 | 53,49,39,43,39 | OK | 20 - 30 | 44,35,26,29,26 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 75,75,89,75 | OK | 40,39,53,39 | OK | 20 - 30 | 27,26,39,27 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 75,72,65,72 | OK | 39,38,30,37 | OK | 20 - 30 | 24,23,16,23 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 79,72,72,72 | OK | 43,39,39,39 | OK | 20 - 30 | 31,26,26,26 | OK | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.06k | OK | 1.0 - 1.1 k | 1.052k | OK | 1.0 - 1.1 k | 1.24k | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 26 | OK | < 1 nS | OK | OK | <1 nS | OK | OK | |
| Qo bias/QET Ret | 11 | 23 | OK | OK | OK | <1 nS | OK | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 24,24 | OK | 6 - 8 nS | 21,21 | OK | <1 nS | 19,19 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 24,24 | OK | 21,21 | OK | <1 nS | 19,19 | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.707 | OK | 0.9 - 1.0 V | 1.084 | OK | — | – | OK |
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.12 | – | 1.2 - 1.5 V | 1.4 | – | — | – | – |
| LED 2 | chassis | 26 | 1.11 | – | 1.392 | – | — | – | – | ||
| Detector: | G106a DIB3 | BH: | Stripline: | 21 | |||||||
| SC: | ZB1-4-8 | 300 K | LN | LHe | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 6 | Date | 2012-12-16 | Date | 2012-12-18 | Date | 2012-12-20 | ||||
| SQUET: | ZC70 | Who | HC & RR | Who | HC | Who | HC | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.9k | OK | 1 - 2 k | 2.0k | OK | 80 - 140 | 100 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.4k | OK | 1.7k | OK | 80 - 140 | 106 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.7k | OK | 1.8k | OK | 80 - 140 | 98 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 5.3k | OK | 2.1k | OK | 80 - 140 | 96 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 29k | OK | 9 - 13 k | 15k | OK | 1.0 - 1.3 k | 1.26k | OK |
| Feedback B/Ret | 2 | 18 | 25k | OK | 9.2k | OK | 1.0 - 1.3 k | 1.26k | OK | ||
| Feedback C/Ret | 41 | 25 | 28k | OK | 14k | OK | 1.0 - 1.3 k | 1.26k | OK | ||
| Feedback D/Ret | 22 | 6 | 35k | OK | 17k | OK | 1.0 - 1.3 k | 1.24k | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 102,102,90,93,90 | OK | 30 - 40 | 63,60,50,54,51 | OK | 20 - 30 | 42,38,28,31,28 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 88,87,100,87 | OK | 44,43,56,44 | OK | 20 - 30 | 28,28,40,28 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 87,85,76,85 | OK | 47,46,37,46 | OK | 20 - 30 | 26,26,17,26 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 96,89,89,89 | OK | 53,49,48,49 | OK | 20 - 30 | 32,27,27,27 | OK | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.066k | OK | 1.0 - 1.1 k | 1.044k | OK | 1.0 - 1.1 k | 1.25k | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | <1 | OK | < 1 nS | open | OK | <1 nS | open | OK | |
| Qo bias/QET Ret | 11 | <1 | OK | open | OK | <1 nS | open | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23,23 | OK | 6 - 8 nS | 20,20 | OK | <1 nS | 18,18 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23,23 | OK | 20,20 | OK | <1 nS | 19,19 | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | open | OK | 0.9 - 1.0 V | 1.091 | OK | — | – | OK |
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.119 | 2 nS | 1.2 - 1.5 V | 1.396 | OK | — | – | – |
| LED 2 | chassis | 26 | 1.108 | open | 1.389 | 23 nS | — | – | 23 nS | ||
| 2005 | Most of the mixture has condensed. Switch to H35 thermometer and start automated circulation procedure. |
|---|---|
| 2040 | Valve V6 on IGH is fully open. Stop automated circulation procedure. Mixing chamber temperature 105 mK and falling. |
| 2230 | Mixing chamber temperature 73.0 mK. Rate of cool down has slowed, but it's still heading in the right direction. Would be interesting to see the situation tomorrow morning. Could potentially flash the detector for a couple of hours, wait for it to cool down, tune the SQUIDs, conduct charge breakdown measurements and measure the Tcs. This is rather ambitious, and also dependent on whether the fridge warms up when the DCRCs are plugged in. There is still cooling power in the fridge (still pressure 0.07 mbar, pressures in G1 & G2 are 76 mbar & 76 mbar respectively). |
| Fill N2 cold trap full. | |
| 2240 | Fed Cricket 4 pellets. |
| Mixing chamber temperature 72.4 mK and falling at same rate as that 10 minutes ago. |