| 0930 | Filled LN2 cold trap. |
| 1020 | Fed Cricket 3 pellets. |
| 1036 | Start automatic circulation process. |
| 1215 | End automatic circulation process. Fridge is stabilizing at about 66 mK. |
| 1229 | Lower 1 K pot temperature to 1.57 K. |
| 1507 | Pumping mixture to dump. Tried plugging in 2 DCRC boards and the fridge warmed up to 81 mK. Ran out of LHe anyway. |
| 1802 | Warmed up fridge with mixing chamber heater. Did a leak check at 10.9 K. |
| | Background leak rate: 1.4E-8 mbar l/s. |
| | IVC rate : 4E-9 mbar l/s. |
| 1820 | Fed Cricket 4 pellets. |
| 2020 | Mixing chamber temperature 15.1 K. Background helium leak rate in hose to IVC 3.6E-9 mbar l/s. Leak rate in IVC shot up to 3.3E-6 mbar l/s. |
| | Evacuate IVC with turbo pump overnight. Leave mixing chamber heater on (power 20 mW). |
| 2125 | Fill N2 cold trap full. |