2012/12/08
| 1205 | Filled LN2 cold trap. |
|---|---|
| Fed Cricket 4 pellets. |
| 1245 | Opened V14 for 5 seconds to add in some more mixture. |
|---|---|
| 1404 | Opened V14 for 5 seconds to add in some more mixture. |
| 1528 | Opened V14 for 7-10 seconds to add in some more mixture. |
| 1530 | Opened V14 for another 3 seconds to add in some more mixture. |
| 1650 | Opened V14 for 10 seconds to add in some more mixture. |
| Time | P1 (mbar) | P2 (mbar) | G1 (mbar) | G2 (mbar) | Mixing Chamber (mK) | Still (kOhms) | Cold Plate (kOhms) | 1 K Pot (K) |
|---|---|---|---|---|---|---|---|---|
| 1230 | 0.06 | 9 | 63 | 64 | 75 | 4.40 | 13.9 | 1.6 |
| 1344 | 0.06 | 9 | 63.8 | 65.9 | 77 | 4.39 | 14.02 | 1.6 |
| 1527 | 0.06 | 12 | 64.5 | 66 | 79.7 | 4.39 | 14.1 | 1.6 |
| 1639 | 0.065 | 9 | 66 | 67.7 | 84.6 | 4.40 | 14.2 | 1.59 |
| 1749 | 0.065 | 9 | 67.6 | 69.3 | 89.7 | 4.41 | 14.4 | 1.58 |
| 1757 | Begin 1-shot test (closed V13a, opened V9). |
|---|---|
| 1912 | The thermometer froze at 37.0005 mK for a couple of minutes. It's functioning again. |
| 1924 | The temperature seems to be increasing again. |
| 1942 | The fridge seems to be experiencing a cooling effect again. |
| 1948 | The temperature seems to be increasing again. (G2 ~ 330 mbar) |
| 2034 | Filled LN2 cold trap. |
| Fed Cricket 4 pellets. | |
| 2041 | Stopped 1 K pot and recovery from 1 K pot. Opened V5. Now pumping mixture to dump from both still and condenser line. |