| 1543 | Mixing chamber temp: 68.6 mK. Will put in DCRC32 now. |
| 1549 | Plugged in DCRC32. Turned on power only by writing 0800. Interestingly, the charge channels seem fine. I didn't turn on the FET heater. |
| | POE reading: 111 W |
| 1553 | Temperature begins to increase dramatically. |
| 1613 | Mixing chamber: 76 mK. |
| 1644 | Temperature has stabilized to 80 mK. Interesting, there was a few minutes delay before this temperature increase. Very similar to what happened when DCRC33 was plugged in on 2012/12/04. |
| 1736 | Filled LN2 cold trap. |
| | Fed Cricket 4 pellets. |
| 1754 | Temperature has gone back down to 68.4 mK. Plugged in DCRC31. |
| | POE reading: 111 W |
| 1757 | This time the charge channels were railed. Had to turn on the FET heater. However, though the channels are no longer railed, they're fluctuating around 2 V. |
| 1846 | Temperature has leveled out to 80.6 mK. Again, similar to the other boards. |