1002 Fed cricket 3 pellets.
1010 Filled LN2 cold trap.
Time P1 (mbar) P2 (mbar) G1 (mbar) G2 (mbar) Mixing Chamber (mK) Still (kOhms) Cold Plate (kOhms) 1 K Pot (K)
1331 0.065 70 77.8 83 72.5 - - 1.74
1341 Reducing 1 K pot temperature to ~ 1.59 K. Forgot to record still and cold plate values before doing so. After this, will add some mixture to circulation.
Time P1 (mbar) P2 (mbar) G1 (mbar) G2 (mbar) Mixing Chamber (mK) Still (kOhms) Cold Plate (kOhms) 1 K Pot (K)
1407 0.055 13 68 74 68.3 4.42 13.7 1.60
1543 Mixing chamber temp: 68.6 mK. Will put in DCRC32 now.
1549 Plugged in DCRC32. Turned on power only by writing 0800. Interestingly, the charge channels seem fine. I didn't turn on the FET heater.
POE reading: 111 W
1553 Temperature begins to increase dramatically.
1613 Mixing chamber: 76 mK.
1644 Temperature has stabilized to 80 mK. Interesting, there was a few minutes delay before this temperature increase. Very similar to what happened when DCRC33 was plugged in on 2012/12/04.
1736 Filled LN2 cold trap.
Fed Cricket 4 pellets.
1754 Temperature has gone back down to 68.4 mK. Plugged in DCRC31.
POE reading: 111 W
1757 This time the charge channels were railed. Had to turn on the FET heater. However, though the channels are no longer railed, they're fluctuating around 2 V.
1846 Temperature has leveled out to 80.6 mK. Again, similar to the other boards.

200

1853 Testing out one of the old PoE's used to power the boards. Let's see if the temperature goes below 80 mK.
As an aside, I didn't have to turn on the FET heater again for DCRC 31. Maybe the channels were still on a high, so to speak. Still oscillating around 2 V.
DCRC31 stayed around 79 mK. Removed board.