UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20121124

2012/11/24

0250 Fed Cricket 5 pellets.
0510 Finished measurements. Helium liquid level 0.08%. Stop 1 K pot, pump mixture out of mixing chamber, turn on mixing chamber heater (power 20 mW), and start warming up.
0525 Reset DCRCs.
0540 Realised I've forgotten to take charge breakdown measurements between charge channels on same side. Helium liquid level is 0.04%, mixing chamber temperature is >7 K, and dump pressure is 490 mbar. There probably isn't enough helium to cool back down again, but it's worth a shot.
Turn off mixing chamber heater, stop pumping mixture out of fridge and restart 1 K pot.
0640 Mixing chamber temperature 2.6 K. Helium liquid level -1.0%, but helium level meter lies a little above 1 K pot intake valve.
Isolate dump and start circulation manually.
1010 Had enough helium to reach >200 mK, but temperature has been rising over last hour or so as helium has run out and 1 K pot has warmed. Mixing chamber temperature currently a little shy of 700 mK. Due to higher base temperature, it is not possible to see phonon pulses. SQUIDs locked regardless.
Charge beakdown measurements for Side 1 indicate that charge breakdown occurs when Qi = +4 V & Qo = 0 V. This is no different from what has been observed at +/-4 V bias, where Qi showed signs of charge breadown. When Qi = 0 V & Qo = +12 V, Qo trace begins to show evidence of charge breakdown. There is no evidence of charge breakdown in the opposite polarity (ie. when Qi = -12 V & Qo = 0 V, or when Qi = 0 V & Qo = -12 V).
For Side 2, Qi shows signs of breakdown when Qi = +4.5 V & Qo = 0 V, and Qi = -8 V & Qo = 0 V. Qo shows signs of breakdown when Qi = 0 V & Qo = +4.5 V, and Qi = 0 V & Qo = -8 V. From previous measurements, Qi trace was all over the place at -/+4 V bias. Cannot seem to repeat this behaviour.
1050 Mixing chamber temperature around 930 mK. Stop 1 K pot, turn on mixing chamber heater (20 mW) and pump mixture into dump. Start warm up procedure.
1230 Dewar lowered, fans started, mixing chamber temperature 25.3 K and rising.
1235 Fed Cricket 3 pellets.
1330 Turn off mixing chamber heater.
1400 Mixing chamber temperature 92 K. Add 5.4 mbar of N2 exchange gas to IVC in an effort to aid warm up.
Run Windows Update on DAQ machine.
1530 Mixing chamber temperature 174 K.
Start evacuating helium transfer stick's vacuum jacket.