UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20121001

2012/10/01

0005 Isolate dump. Pressure of mixture remaining in dump 379 mbar.
0025 Valve V12A on IGH fully open. Have been adjusting needle valve during condensation procedure: temperature of 1 K pot 1.88 K. Mixing chamber temperature 5.9 K and falling.
0240 Mixing chamber temperature 2.6 K. Reduce 1 K pot temperature to 1.76 K.
0305 Mixing chamber temperature 2.2 K. Reduce 1 K pot temperature to 1.71 K.
Fill N2 cold trap full.
0310 G1 & G2 pressures are between 60 and 70 mbar. Most of the mixture has condensed. Start automated circulation procedure.
0450 Mixing chamber temperature 63.8 mK. Stop automated circulation procedure. Pressures in P1, G1 & G2 are 0.06 mbar, 51 mbar & 60 mbar respectively. Rate of cooling has slowed down, but temperature still seems to be decreasing. The still pressure appears to indicate that it may be running dry. Adding more mixture may help. It may be worth being greedy and playing around with settings to see if we can drive the temperature lower. Good news so far though - let's hope it lasts.
1322 Cursory electrical check
Dib 1
SQUID bias B/Ret : 99 Ohms
Qi bias/QET Ret: < 1 nS
Dib 2
Qi bias/QET Ret: < 1 nS
Qo bias/QET Ret: < 1 nS
1330 Filled LN2 cold trap.
1656 Still: 4.47 kOhms ; Cold plate: 14.6 kOhms
1930 Plug DCRC33 into DIB 1. Temperature increases a little (~1 mK) before falling again.
2010 Plug DCRC32 into DIB 2. Temperature increases a little (~1 mK) before falling again.
Base temperature appears to have stabilised at 62.5 mK.
2340 Start flashing detector: LED 2 on DIB1; and LEDs 1 & 2 on DIB2. Usual settings.