2012/10/01
| 0005 | Isolate dump. Pressure of mixture remaining in dump 379 mbar. |
|---|---|
| 0025 | Valve V12A on IGH fully open. Have been adjusting needle valve during condensation procedure: temperature of 1 K pot 1.88 K. Mixing chamber temperature 5.9 K and falling. |
| 0240 | Mixing chamber temperature 2.6 K. Reduce 1 K pot temperature to 1.76 K. |
| 0305 | Mixing chamber temperature 2.2 K. Reduce 1 K pot temperature to 1.71 K. |
| Fill N2 cold trap full. | |
| 0310 | G1 & G2 pressures are between 60 and 70 mbar. Most of the mixture has condensed. Start automated circulation procedure. |
| 0450 | Mixing chamber temperature 63.8 mK. Stop automated circulation procedure. Pressures in P1, G1 & G2 are 0.06 mbar, 51 mbar & 60 mbar respectively. Rate of cooling has slowed down, but temperature still seems to be decreasing. The still pressure appears to indicate that it may be running dry. Adding more mixture may help. It may be worth being greedy and playing around with settings to see if we can drive the temperature lower. Good news so far though - let's hope it lasts. |
| 1322 | Cursory electrical check |
| Dib 1 SQUID bias B/Ret : 99 Ohms Qi bias/QET Ret: < 1 nS |
|
| Dib 2 Qi bias/QET Ret: < 1 nS Qo bias/QET Ret: < 1 nS |
|
| 1330 | Filled LN2 cold trap. |
| 1656 | Still: 4.47 kOhms ; Cold plate: 14.6 kOhms |
| 1930 | Plug DCRC33 into DIB 1. Temperature increases a little (~1 mK) before falling again. |
| 2010 | Plug DCRC32 into DIB 2. Temperature increases a little (~1 mK) before falling again. |
| Base temperature appears to have stabilised at 62.5 mK. | |
| 2340 | Start flashing detector: LED 2 on DIB1; and LEDs 1 & 2 on DIB2. Usual settings. |