2012/09/05

0955 Fed Cricket 3 pellets.
1010 Filled LN2 cold trap.
1340 Mixing chamber temperature 7.05 mK and appears to be holding steady. Close all valves to mixing chamber lines in fridge. Helium leak rate in pipes leading to IVC 1.4E-9 mbar l/s. Open up IVC and helium leak rate rises to 2.2E-6 mbar l/s. It is doubtful there is a superfluid leak.
Isolate IVC and leak checker. Use turbo pump to evacuate pipes to IVC.
1400 Start helium transfer. Helium liquid level -0.1%.
1445 End LHe transfer (88.4%).
1630 Fed Cricket 4 pellets.
2340 Fill N2 cold trap full.
Uncalibrated He-3 leak rate in pipes leading to IVC 1.8E-10 mbar l/s. Open up IVC and uncalibrated He-3 leak rate falls to 1.7E-10 mbar l/s. Let's see what things look like in the morning, but it appears that there isn't any trace of He-3 which would rule out a leak from the dilution unit.
2355 Start cryocleaning mixture through nitrogen cold trap 2.