| 1515 | Fill N2 cold trap full. |
| | Fed Cricket 4 pellets. |
| 1540 | Mixing chamber temperature 6.98 K, and hovering around the same temperature. Helium leak level in pipes leading to IVC 1.5E-9 mbar l/s. Open up IVC and helium leak rate increases to 1.5E-6 mbar l/s. Value is fairly constant. Looks like 1 K pot is warm enough and we can proceed with leak checking test. |
| 1545 | Isolate IVC and start 1 K pot: temperature 1.6 K. |
| | Isolate leak checker and pump on lines to IVC with turbo pump. |
| 1920 | Temperature of 1 K pot has been 1.76 K for some time. Stop 1 K pot. |