| 1510 | Run regeneration cycle again. |
| 1615 | Run regeneration cycle again. |
| 1735 | Run regeneration cycle again. |
| 2330 | Pumped mixture into dump, sealed it off, and cryopumped mixture from exhaust into cryopump. Cryopump was sealed off and left in liquid helium dewar. |
| | Sealed off fridge. Turned off and drained oil from He-3 pump. Oil was discoloured and smelt burnt. Ran pump for additional 10 seconds and drained remaining oil. Filled pump with 3 l of oil and turned it on. |
| | Cryocleaned He-3 pump exhaust through nitrogen cold trap 2 and valve 3, bypassing dump, for 10 minutes. Opened dump to He-3 pump exhaust and pumped mixture from cryopump to exhaust with He-3 pump. |
| 2350 | Mixing chamber temperature has oscillated around 6.07 K since 1405. It seems to indicate that there is no leak. Would we expect the temperature to rise due to heat load from conduction? |
| | Conducted quick helium leak check: 4.6E-6 mbar l/s, which would indicate that there is a leak. Could this be due to the large helium background in the room from cryopumping the mixture? |