| 0950 | Start flashing the detector for 1 hour. |
| 1010 | Fed Cricket 3 pellets. |
| 1105 | Filled LN2 cold trap. |
| 1303 | Need to clean the cold traps. Will leave mixture that's in the fridge there and pump the rest to the dump. Close condenser line speedivalve, V6, V13 A/B. Open V12A/B, 9, 5. |
| 1436 | Removed DCRC boards. Pulled out cold traps. Pumping on cold traps with He-4 pump, waiting for them to warm up to room temperature. |
| 1700 | Fed Cricket 4 pellets. |
| 1708 | Heated cold traps. Entered cold traps back into respective baths. Helium cold trap not able to go back in all the way. |
| 1730 | Re-started 1K pot |
| 1751 | Begin LHe transfer (8.4%). |
| 1845 | End LHe transfer (91.9%). |
| | Filled LN2 cold trap. |
| 1911 | Re-commence circulation. |
| 1954 | The temperature dropping rate significantly decreased around MC = 103mK. G2=57mbar, G1=47mbar. There was still a large pressure difference across the N2 cold trap. |
| 2002 | Added N2 cold trap 2 into circulation. The pressure difference of G2 and G1 reduced and the temperature dropped faster. |
| 2014 | MC=92.8mK. Temperature dropping rate starts to decrease again. G2=42mbar, G1=45mbar, P1=0.045mbar. Still=4.55kOhm. |
| 2017 | Add 2mW heating power to still. P1 goes up to 0.075mbar. Still=4.47kOhm. G2=44.8mbar, G1=49mbar. |
| 2022 | MC=89.5mK and cools very slowly. Turn off the still heater. The He cold trap may still have a large impedance. |
| 2028 | Unplug all DCRC's in case they are heating up the fridge. No obvious effect observed MC=89.5mK. |
| 2136 | DCRC33 refuse to power up. Not even to take charge data. I'll flash the detectors. |
| 2141 | Flash detector with LED1 on DIB3/DCRC31, LED1&2 on DIB2/DCRC32. ON=100us, OFF=5000us, current=-0.6mA. |