UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20120706

2012/07/06

0030 Fill N2 cold trap full.
1005 Filled LN2 cold trap.
1220 Mixing chamber temperature 89.7 mK. Let's try to isolate the fridge. Unplug pumps and detach dewar from pulley system.
1415 Well that did nothing… turn on still heater again (1 mW) and wait. See what happens.
1610 Mixing chamber temperature has increased to 99.8 mK. Pressures of G1, G2 & P1 are 54 mbar, 62 mbar & 0.08 mbar respectively. Other temperature stages:
Table 1: Temperatures of stages when still heater powered to 1 mW.
Stage Measured Resistance [kΩ]
Still 4.46
Cold plate 14.9
Mixing chamber 20.8
1615 Increase still heater power to 2 mW.
1800 Mixing chamber temperature has increased to 109.6 mK. Pressures of G1, G2 & P1 are 56 mbar, 69 mbar & 0.095 mbar respectively. Other temperature stages:
Table 2: Temperature of stages when still heater powered to 2 mW.
Stage Measured Resistance [kΩ]
Still 4.41
Cold plate 14.9
Mixing chamber 19.1
1805 Increase still heater power to 3 mW.
1905 Mixing chamber temperature has increased to 116.8 mK. Pressures of G1, G2 & P1 are 60 mbar, 73 mbar & 0.11 mbar respectively. Other temperature stages:
Table 3: Temperature of stages when still heater powered to 3 mW.
Stage Measured Resistance [kΩ]
Still 4.38
Cold plate 14.9
Mixing chamber 18.0
1915 Increase still heater power to 4 mW.
1945 Start helium transfer. Helium liquid level 25.7%.
2036 End LHe transfer (91.3%)
2110 Mixing chamber temperature has increased to 119.0 mK. Slight decrease from peak temperature witnessed (<1 mK), which may have to do with helium transfer. Pressures of G1, G2 & P1 are 71 mbar, 82 mbar & 0.12 mbar respectively. Other temperature stages:
Table 4: Temperature of stages when still heater powered to 4 mW.
Stage Measured Resistance [kΩ]
Still 4.38
Cold plate 15.0
Mixing chamber 17.8
2120 Increase still heater power to 5 mW.
2305 Mixing chamber temperature has increased to 120.8 mK. Pressures of G1, G2 & P1 are 79 mbar, 87 mbar & 0.13 mbar respectively. Other temperature stages:
Table 5: Temperature of stages when still heater powered to 5 mW.
Stage Measured Resistance [kΩ]
Still 4.37
Cold plate 14.9
Mixing chamber 17.6
2310 Turn still heater off.
2335 Fill N2 cold trap full.