2012/05/07
| 1100 | Filled LN2 cold trap. |
|---|---|
| 1630 | Start tower assembly procedure. |
| 1800 | Conduct detector level continuity checks: |
| Table 1: Detector level continuity checks | ||
|---|---|---|
| DIB 1 | DIB 2 | |
| Signal | Measured | Measured |
| QET D | 220 Ω | 235 Ω |
| LED1 R forward | 765 kΩ | 783 kΩ |
| LED1 V forward | 1.060 V | 1.060 V |
| LED1 G backward | <0.7 nS | <0.1 nS |
| LED2 R forward | - | 766 kΩ |
| LED2 V forward | - | 1.060 V |
| LED2 G backward | - | <0.4 nS |
| Qo(+ve)-Qi(-ve) | 432 Ω | 423 Ω |
| Qi(+ve)-Qo(-ve) | 240 Ω | 273 Ω |
| 1900 | Conduct tower level continuity check. |
|---|
| Detector | G102 | DIB | 1 | |
|---|---|---|---|---|
| side coax | ZB1-1-6 | date/who: | 2012-05-07 | HC & SM |
| face | 2 | |||
| +ve lead | -ve lead | signal | measured | |
| 7 | 8 | QET D | 311 Ω | |
| 9 | chassis | LED1 Rforward | 764 kΩ | |
| 10 | chassis | LED2 Rforward | - | |
| 9 | chassis | LED1 Vforward | 1.100 V | |
| 10 | chassis | LED2 Vforward | - | |
| chassis | 9 | LED1 Greverse | <0.7 nS | |
| chassis | 10 | LED2 Greverse | - | |
| 7 | 11 | Qo Rbias | 26.0 nS | |
| 12 | 11 | Qo C coupling | <0.1 nS | |
| 12 | 13 | Qo Rfeedback | 26.0 nS | |
| 7 | 14 | Qi Rbias | 26.0 nS | |
| 15 | 14 | Qi C coupling | <0.1 nS | |
| 15 | 16 | Qi Rfeedback | 26.0 nS | |
| Detector | G102 | DIB | 2 | |
|---|---|---|---|---|
| side coax | ZB1-1-9 | date/who: | 2012-05-07 | HC & SM |
| face | 5 | |||
| +ve lead | -ve lead | signal | measured | |
| 7 | 8 | QET D | 325 Ω | |
| 9 | chassis | LED1 Rforward | 782 kΩ | |
| 10 | chassis | LED2 Rforward | 768 kΩ | |
| 9 | chassis | LED1 Vforward | 1.100 V | |
| 10 | chassis | LED2 Vforward | 1.100 V | |
| chassis | 9 | LED1 Greverse | <0.1 nS | |
| chassis | 10 | LED2 Greverse | <0.5 nS | |
| 7 | 11 | Qo Rbias | 25.0 nS | |
| 12 | 11 | Qo C coupling | <0.1 nS | |
| 12 | 13 | Qo Rfeedback | 25.0 nS | |
| 7 | 14 | Qi Rbias | 26.0 nS | |
| 15 | 14 | Qi C coupling | <0.1 nS | |
| 15 | 16 | Qi Rfeedback | 26.0 nS | |
| 1930 | Tower assembly procedure completed. |
|---|---|
| 2315 | Filled LN2 cold trap. |