0030 Pump out exchange gas at 8 K.
Helium transfer complete. Helium liquid level 92.7%.
Fill N2 cold trap full.
1330 Mixing chamber temperature 5.4 K. Looks good; no sign of any leak. Start condensing mixture.
1600 Mixture condensed relatively slowly. Mixing chamber temperature 3.4 K. Start circulating mixture.
1700 Stop automated circulating procedure and run manually. Bleed mixture from dump into nitrogen cold trap through valve V14.
1730 Tower temperature has dropped below 100 mK. Stop bleeding mixture. We're ready to take data!
2200 Encountered a big problem. Plugged DCRC33 into flexible side coax and Qi and Qo signals saturated when 15 V power supply to FET was turned on. Switched board with DCRC32. Saw a decent PSD, and no amount of hammering seemed to have any effect. Connected DCRC33 to vacuum side coax, cycled the power and set the timing link: PSDs were fine on vacuum side coax; but Qi & Qo signals on flexible side coax saturated. Turned on FET heater for 5 minutes, broke timing link, nothing worked. Electrical check with breakout box shows no problems with FET heater. It's a bit of mystery. Email sent around to ask for suggestions.