| 0845 | IVC pressure 4e-3mbar. There is a big leak somewhere. Need to find it out. |
| 1253 | Leak check - background rate is 2.3E-9 mbar l/s. |
| | We sprayed helium everywhere and there was no leak detected. The pressure gauge is a suspect here. It's reading the same pressure as yesterday when we though there was a leak and after the leak tests and spraying helium the pressure reading didn't waver. |
| | The IVC was isolated and the pressure did not decrease. It stayed at 4E-3 mbar. |
| | We're going to go ahead and raise the dewar and do another leak check. |
| 1514 | Raised main bath. Pumping on main bath now. |
| 1537 | Filled main bath with He gas. Cleaning mixture. Background leak rate before sealing off leak checker from IVC was : 4.2E-9 mbar l /s |
| 1557 | Filled IVC line with 14 mbar of N2 exchange gas. When opening the valve into the IVC, pressure dropped to 1.6 mbar. |
| 1610 | Begin LN2 transfer. |
| 1900 | End LN2 transfer. |