UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20120223

2012/02/23

0950 Filled LN2 cold trap.
1051 Data list from last night was finished. The vi didn't stop flashing after the data was done. Kind of a bug. The temperature wasn't as stable as expected but still within 3mK.
# -10V, 2K, Am-241 & Ba-133, data/flash/cooldown = 120/3/5 min. flashed 35min before data
4.0G    07120222_2303
# +10V, 2K, Am-241 & Ba-133, data/flash/cooldown = 120/3/5 min.
5.9G    07120223_0112
# -8V, 2K, Am-241 & Ba-133, data/flash/cooldown = 30/3/5 min.
1.5G    07120223_0321
1.5G    07120223_0400
1.5G    07120223_0438
1.5G    07120223_0516
# +8V, 2K, Am-241 & Ba-133, data/flash/cooldown = 30/3/5 min.
1.5G    07120223_0554
1.5G    07120223_0633
1.5G    07120223_0712
1.5G    07120223_0750
# -6V, 2K, Am-241 & Ba-133, data/flash/cooldown = 30/3/5 min.
1.5G    07120223_0828
1.5G    07120223_0906
1.5G    07120223_0944
1.4G    07120223_1023
Temperature during data taking
1149 Turn off LS370 heater. Close V3. Open V6 slowly. Start circulation again.
1955 Pulled out He cold trap. G1 went up to 160mbar then dropped.
2001 Start pumping both the N2 cold trap and He trap while warming up.