2012/02/21
| 0040 | Taking 07120221_0040_F0001. First of four series at +7V Am-241 data. Data/flash/cooldown = 30/0.1/20 min. Flashed 1 hour before taking data. Trigger on Q1-4 at 40 ADC bins. Will take +10V for four series, +5V two series, and -5V two series. Tower temperature 80mK. |
|---|---|
| 0945 | Filled LN2 cold trap. |
| 1145 | Data was finished. Raise temperature to 600mK. |
| 1212 | 07120221_1212_F0001. -10V Am-241 data at 600mK. |
# 1 hour flashing before the data below # +7V Am-241 Data/flash/cooldown = 30/0.1/20 min 80mK 1.7G 07120221_0040 1.7G 07120221_0130 1.7G 07120221_0220 1.7G 07120221_0310 # +10V Am-241 Data/flash/cooldown = 30/0.1/20 min 80mK 1.7G 07120221_0401 1.7G 07120221_0452 1.7G 07120221_0542 1.7G 07120221_0632 # +5V Am-241 Data/flash/cooldown = 30/0.1/20 min 80mK 1.7G 07120221_0723 1.7G 07120221_0813 # -5V Am-241 Data/flash/cooldown = 30/0.1/20 min 80mK 1.7G 07120221_0904 1.6G 07120221_0954 # -10V Am-241 Data/flash/cooldown = 30/0.1/20 min 600mK 3.0G 07120221_1212
| 1720 | Plug in both HV DCRC interface boards and apply negative bias to “ground” electrode (ie. collect electrons). Signs of charge breakdown first seen at -40 V. Complete charge breakdown at -45 V. May be able to take data at -35 V. |
|---|---|
| 1820 | Start flashing detector. |
| 1833 | Start LHe transfer (11.6%). |
| 1915 | Finish LHe transfer (92.8%). |
| Stop flashing detector. | |
| 1929 | Filled LN2 cold trap. |
| 1930 | Close V6 and open V14. Stop circulation. Set temperature to 1.5K. |
| 1938 | Change PID from (10, 120, 0) to (10, 10, 0). |
| 1946 | 1.5K is very close to 1K pot temperature. Set temperature to 2K to get a larger temperature gradient. |
| 1950 | Close V14, Open V3 and V9 instead so the mixture has to go through the N2 cold trap. |
| 1956 | Change PID from (10, 10, 0) to (10, 60, 0) based on the temperature oscillation period. |
| 2002 | Temperature overshoots frequently. Change PID from (10, 60, 0) to (5, 60, 0). |
| 2008 | T still overshoots. Set PID to (2, 0, 0) |
| 2010 | Change IDR heater range to 2mW from 20mW. |
| 2013 | T is well below the setting point. Set PID to (10, 0, 0). |
| 2055 | Checked U02844 calibration data. At 2K excitation is 2mV rather than 20uV. Now the temperature reading is much more stable. |
| 2101 | Using PID (1000, 0, 0), temperature stabilizes at 1.9992K. It good enough to take data. I'll stop the tuning here though could use a better I. |
| Final settings: IDR MC heater range 2mW; LS370 PID (1000, 0, 0), heater range 10mA; and IGH settings below. |
| 2111 | Temperature rose to 1.9994K slowly. Still good enough to take data. |
|---|---|
| 2130 | Tried 5min data and 5min flashing at 2K. Can see clear pulses at 0V but the peak the pulses are rounded a lot. The temperature kept increasing at roughly a constant rate up to 2.07K then dropped when the flashing was done. The cooldown took 6min. |
| 2150 | 3min flashing took 4min to cooldown. |
|---|---|
| 2208 | Start taking data 07120221_2206_F0001. -10V bias at 2K with Am-241 source. Data/flash/cooldown = 30/3/5 min. Trigger on Q1-4 at 40ADc bin. Will take 4 series for the bias. Will take data at +10, -8, +8, -6V with the same DAQ settings. |