Data collected in the last run
# -7V Am-241 2 hour long run. Trigger on Q1-4 at 40 ADC bin.
4.2G ./07120217_2247
# 0V data/flash/cooldown = 120/0.1/20 min
4.9G ./07120218_0048
# +7V Am-241 2 hour long run. Trigger on Q1-4 at 40 ADC bin.
4.6G ./07120218_0313
# 0V data/flash/cooldown = 120/0.1/20 min
4.9G ./07120218_0514
# -10V Am-241 2 hour long run. Trigger on Q1-4 at 40 ADC bin.
4.6G ./07120218_0739
| | There is a temperature peak from 0750 to 0915 up to 106mK from 80mK. Otherwise stable. |
| 1112 | Take 0V data 07120218_1112_F0001. Continue the list. T=78mK. |
| 1313 | Filled N2 cold trap. |
| 1740 | All data in the last list finished. |
Data collected in the last run
# 0V data/flash/cooldown = 120/0.1/20 min
4.9G ./07120218_1112
# +10V Am-241 2 hour long run. Trigger on Q1-4 at 40 ADC bin.
4.6G ./07120218_1337
# 0V data/flash/cooldown = 120/0.1/20 min
5.0G ./07120218_1538
| 1748 | Raise fridge temperature to 400mK. |
| 1758 | Fridge is at 400mK. Try taking data 1min and flash 0.1min and see how long it takes tool cool to 400mK. |
| 1805 | It took 2min to cool down to 400mK. |
| 1847 | The event rates are low in the recent data sets. The detector may need to be neutralized. Flash the detector for 2min and see how long it takes to cool down to 400mK. |
| 1911 | It took 5min to cool down to 400mK for longer flashing. |
| 1919 | Labview crashed. |
| 1940 | Start flashing detector with Ba-133 source for 12 hours and then cooldown 10min to 400mK. Then take Ba-133 data at -10, +10, -8, +8, -6, +6, -4, +4, -2, +2V biases on Q1-4. Data/flash/cooldown = 15/0.1/2 min. 4 series at each bias. |