| 1627 | Tried turning up the voltage of the Kepco power supply, tower temperature jumped at around 20V, charge signal baseline waggled too. Reduced bias to 15V. Temperature dropped and charge baseline was getting stable. |
| 1639 | Tower temperature rose to 81.7mK at 15V bias from 78mK at 0V bias. Probably we will not able to use a high voltage. |
| 1640 | Turn off Kepco power supply. Temperature starts to drop. |
| 1709 | Tower temperature 78.7mK. Set Kepco to 20V and then turn on the power supply. The charge noise jumped up and tower temperature spiked. Does the crystal break down at 20V? |
| 1715 | Removed the Cu tape between the Al case and ebox, the noise got slightly worse. |
| 1815 | Had a look at the synchronization of the boards with different trigger thresholds. It's easier to see if they are synchronized by diabling the trigger from one board. |
| 1843 | Begin LHe transfer (18%). |
| 2207 | Tower temperature 123 mK and keeps fluctuating. Is due to the hi pressure in the bath? |